wirelessSensorPlot

Legacy software

wirelessSensorPlot is no longer maintained by i4M technologies. It has been replaced by nemi Studio (Windows desktop application) and the nemi Studio Web App (browser application). This page is kept for reference only. It describes wirelessSensorPlot up to version 7.4.

Main window of wirelessSensorPlot with three nemi G+ sensor nodes transmitting live data


Product Description

wirelessSensorPlot is a Windows PC application for the configuration, live display and CSV data logging of nemione wireless sensors and telemetry modules (e.g. nemi G+, nemi DAQ) as well as OEM-specific i4M wireless sensors in the nemi Link 2400 radio network. It serves as the primary human-machine interface for the nemi Link 2400 radio ecosystem, allowing engineers to commission sensor networks, monitor measurement quality in real time, and record data for post-processing — all without writing code.

The software communicates with wireless sensor nodes (nemi G+, nemi G+ nano, nemi G+ ultra, nemi G+ ultra nano, nemi DAQ, nemi DAQ nano) through one of two receiver types: the nemi Connect USB receiver (direct USB connection to the PC) or the nemi EdgeBase network gateway (TCP/IP connection over LAN or VPN, available from wirelessSensorPlot v7.3 onward). It does not replace the nemi EdgeBase's autonomous logging capabilities; rather, it complements them by providing live visibility and manual control.

Within the product family, wirelessSensorPlot occupies the PC-software layer above the receivers. Hardware configuration of sensor node radio parameters (channel, radio mode, node number) is performed separately by nemiSensorConfig. Once nodes are configured, wirelessSensorPlot handles everything at runtime: sending measurement settings to nodes, displaying live plots, managing recording sessions, and exporting CSV files for analysis in external tools.

The application is distributed as a ZIP archive and requires no installer. It runs on Windows 10 or higher; versions for Linux or macOS were available on request. A single executable supports all sensor types and receiver configurations.


Typical Applications

Where wirelessSensorPlot excels

  • Field commissioning and verification — connecting to an already-deployed sensor network to confirm reception quality, check sample rates, and validate measurement ranges before a longer unattended logging campaign.
  • Lab and bench testing — running a nemi Connect directly on a laptop to characterize sensor behavior, compare ranges, and generate reference CSV files for algorithm development.
  • Live remote access to deployed EdgeBase systems — using wirelessSensorPlot v7.3 or later to connect over VPN to an EdgeBase in the field, without being physically present at the installation.
  • Short-duration measurement campaigns where the PC remains connected throughout, recording CSV files locally.
  • System diagnostics — checking signal strength, battery voltage, and sample-loss rates to troubleshoot radio or sensor issues before contacting support.

Applications by domain

Application Notes
Structural vibration monitoring (bridges, cables, hangers) Typical setup with nemi G+ and nemi Connect or EdgeBase
Rotating machinery and bearing monitoring Check gyroscope range for high RPM; use correct acceleration range
Wind energy (rotor blades, nacelles, towers) Use 500 kbit/s or 125 kbit/s radio mode for RF-challenged environments
Strain gauge measurements (DMS) nemi DAQ with mV/V input; use High-Res mode to avoid spikes
Temperature logging (Pt100/Pt1000) Requires nemi DAQ with constant-current configuration; automatic unit conversion

Technical Specifications

System requirements

Parameter Value
Operating system Windows 10 or higher
Installation No installer; run wirelessSensorPlot_7_x.exe directly from extracted ZIP
Permissions Standard user; depending on the system settings, administrator rights may be required to log files
Receiver interfaces USB (nemi Connect, virtual COM port); TCP/IP (nemi EdgeBase, port 9413)

Supported sensor nodes

Sensor node Type Notes
nemi G+ Vibration / acceleration M08 (8 g) or M40 (40 g) variant; primary MEMS + IMU. The M256 variant is supported only by nemi Studio 1.1 or later
nemi G+ nano Vibration / acceleration Compact form factor; same software settings as G+
nemi G+ ultra High-speed acceleration Up to 26.6 kHz; requires wirelessSensorPlot v7.4 b3 or later
nemi G+ ultra nano High-speed acceleration Same software as G+ ultra
nemi DAQ Telemetry (analog + digital) Up to 3 analog channels, up to 6 digital inputs (DIN), IMU
nemi DAQ nano Telemetry (analog + digital) Compact OEM board; same software settings as nemi DAQ
Parameter Value
Frequency band 2.4 GHz ISM
Channel range 0–79
Radio modes 2 Mbit/s, 1 Mbit/s, 500 kbit/s, 125 kbit/s
Max nodes per channel 1, 3, or 8
Node numbers 1–8 (must be unique per channel)

Aggregate sample rate capacity and per-node rates (nemi G+):

Radio mode Max. aggregate Hz 3 nodes 8 nodes 1 node (low latency)
2 Mbit/s 36,000 Hz 4,000 Hz/node 1,000 Hz/node 4,000 Hz/node
1 Mbit/s 15,000 Hz 2,000 Hz/node 500 Hz/node 4,000 Hz/node
500 kbit/s 8,000 Hz 1,000 Hz/node 250 Hz/node 2,000 Hz/node (4,000 Hz marginal)
125 kbit/s 3,000 Hz 250 Hz/node 125 Hz/node 1,000 Hz/node

Per-node values for nemi G+ (three axes) with a radio channel load below 70 % according to the nemi Link 2400 Radio Channel Calculator.

Wireless range (approximate, see nemi Link 2400):

Radio mode In buildings Urban environment
2 Mbit/s ~20 m ~50 m
1 Mbit/s ~50 m ~125 m
500 kbit/s ~90 m ~225 m
125 kbit/s ~140 m ~350 m

Reference distances of the Signal Strength indicator (since v7.0): the received RSSI value is converted into a distance and shown relative to the maximum distance defined for the selected radio mode (linear scale). These are theoretical line-of-sight values under optimal conditions, not typical ranges:

Radio mode Max. distance (line of sight, optimal conditions) Link budget
2 Mbit/s 100 m 94 dB
1 Mbit/s 250 m 102 dB
500 kbit/s 450 m 107 dB
125 kbit/s 700 m 111 dB

Data output

Parameter Value
Format CSV (one or more files per sensor slot per recording session)
Delimiter Semicolon (;)
Decimal separator Period (.)
Timestamps Three columns per row: Timestamp (ASCII), Timestamp (s) (UNIX epoch, µs resolution), Timecounter (s)
Unit conversion Applied before write, including slope a and offset b; raw ADC codes are not stored

EdgeBase IMU (when using EdgeBase as receiver)

Select EdgeBase IMU in the node drop-down menu of any tab; the data is shown in the lower IMU plot of that tab. The settings are fixed by wirelessSensorPlot and cannot be changed; the EdgeBase IMU temperature is not displayed:

Parameter Value
Sample rate 416 Hz
Accelerometer range ±4 g
Gyroscope range ±500 °/s
Magnetometer range ±8 Gauss

First Commissioning

Prerequisites

  • PC with Windows 10 or higher
  • wirelessSensorPlot ZIP extracted to a local folder
  • One or more sensor nodes with radio parameters already configured via nemiSensorConfig
  • A nemi Connect USB receiver or a reachable nemi EdgeBase (with logging_mode: 6)

Step 1 — Configure sensor nodes (one-time per node)

Radio parameters (channel, node number, max nodes per channel, radio mode) are stored permanently on the sensor node and must be set once via nemiSensorConfig before the first use.

  1. Turn the sensor node off.
  2. Connect it to the PC via USB. Use a data-capable USB cable (not a charge-only cable).
  3. Open nemiSensorConfig.
  4. Turn the sensor node on — it shows solid green + flashing yellow and should appear in the software.
  5. Click Connect.
  6. Wait for:

Waiting for RF Parameters / Config Message to (re)start ...
7. Set: radio mode, max nodes per channel, channel, and a unique node number. 8. Click Set nemi Link Parameters. 9. Disconnect, close nemiSensorConfig.

Timeout warning: If the node sits connected via USB for more than ~10 seconds without nemiSensorConfig completing a connection, it exits USB configuration mode and switches to radio mode. If you see no response: power-cycle the node, close and reopen nemiSensorConfig, and retry.

Every node on the same channel must have a unique node number. All nodes and the receiver must be set to the same channel, radio mode, and max-nodes count. Mismatches result in no data.

Step 2 — Connect via nemi Connect (USB)

  1. Disconnect the nemi Connect from the PC.
  2. Start wirelessSensorPlot.
  3. Plug the nemi Connect into USB. Windows plays the "new USB device" sound; the LED on the nemi Connect blinks slowly red.
  4. wirelessSensorPlot auto-detects the COM port. If not, click Refresh COM-Ports.
  5. Click Connect COM-Port.
  6. Wait until the system console (top right) shows:

Waiting for Config Messages to start ...
7. (Optional) Load a previously saved configuration: Load Settings from File (optional) → select a *.ini file. 8. (Optional) Restore defaults: Reset Settings (DAQ) sets all nodes to nemi DAQ / nemi DAQ nano, Reset Settings (G+) sets all nodes to nemi G+ / nemi G+ nano. For mixed networks configure the nodes individually and save the settings with Save Settings to File. 9. Select radio data rate and max nodes per channel — these must match the sensor node configuration (set with nemiSensorConfig). 10. Set the radio channel. 11. Set sensor type and measurement parameters for each node, unless they were loaded from a *.ini file. 12. Click Start Receiver. 13. Switch on the sensor nodes on the configured channel. Live data appears in the plots and the nemi Connect LED flashes green.

Radio settings note: The radio mode and max-nodes count can only be changed after clicking Disconnect COM-Port. To change only the channel while the receiver is running, use the Set Channel button.

Step 3 — Connect via nemi EdgeBase (network, v7.3+)

The nemi EdgeBase must be configured with logging_mode: 6 in its sensor_config.yml. It then provides all measurement data live on TCP/IP port 9413, and wirelessSensorPlot accesses it like a nemi Connect — all functions including logging are available.

  1. Start wirelessSensorPlot v7.3 or later.
  2. Click Connect EdgeBase.
  3. Enter the IP address of the EdgeBase. Leave the port at 9413. The PC must be able to reach the EdgeBase via VPN or the local network.

Dialog for connecting a nemi EdgeBase

  1. Click Connect to nemi EdgeBase.
  2. Wait until Waiting for Config Messages to start ... appears as the last message in the system console.
  3. Configure radio network and node parameters (same as with nemi Connect).
  4. Click Start Receiver.
  5. To disconnect: click Disconnect EdgeBase.

LAN vs. LTE: If the EdgeBase has both LAN and LTE available, it prefers LAN. If LAN is unavailable it switches to LTE and stays there until restarted.

VPN connectivity: Set lte_check_ip: 8.8.8.8 in the EdgeBase system_config.yml. A different or unreachable address can cause the EdgeBase to incorrectly assume its LTE connection has failed and restart it.

Step 4 — Start a recording

After the receiver is running and data appears in the plots:

  1. Wait 10 seconds after connecting to the receiver before starting a recording. wirelessSensorPlot uses this time to synchronize the sensor clocks with the PC clock. To synchronize the PC clock itself to another clock, configure a time server (NTP) in Windows.
  2. Select the options below Start Recording if required: Log full Timestamps, Log Debug only, Record Sync.
  3. Click Start Recording and choose an output folder. The CSV files are written in real time; the file names contain date and time of the recording start and, since v7.2, the node names.
  4. Click Stop Recording to end the session.

Files written per sensor slot (x = slot number; the first line contains the column names):

File pattern Content
<datetime>_imu_data_x.csv Integrated MEMS IMU of the node (ACC / GYR / MAG), temperature, battery voltage, RSSI
<datetime>_vibration_data_x.csv Wireless acceleration sensors (e.g. nemi G+): primary MEMS accelerometer in g, including slope a and offset b
<datetime>_telemetry_data_x.csv Telemetry modules (e.g. nemi DAQ): analog channels in the selected unit, including slope a and offset b
<datetime>_din_data_x.csv Telemetry modules (e.g. nemi DAQ): digital inputs
<datetime>_timing_data_x.csv Time synchronization data (only when Record Sync is enabled)
<datetime>_debug_x.csv Debug data (content of all consoles); can be sent to i4M for troubleshooting

Each sample has three time columns: Timestamp (ASCII) (date, time, microseconds), Timestamp (s) (UNIX time, µs resolution) and Timecounter (s) (running time, set to 0 at the start of the recording).

LED status codes (nemi Connect)

LED behavior Meaning
Slow red blink Connected to PC, waiting for configuration
Green steady / green blink Active, receiving sensor data
Rapid / random blinking on sensor node Sensor cannot find a receiver on its configured channel
Green blink every ~10 s (sensor node) Sensor is in sleep mode

Status indicators in wirelessSensorPlot

Background color Meaning
Green All data streams complete; no missing samples
Yellow Last received sample is older than 2 seconds. This is not yet a data loss: with poor reception the sensors buffer data and deliver it later. A sensor in sleep mode also shows yellow
Red At least one sample lost within the last second — check the reception and reduce the sampling rate if necessary

Further displays:

Display Meaning
Battery voltage Above the upper plot of a tab; integrated Li-Po battery: 4.1 V ≈ 100 % charge, approx. 3.3 V ≈ below 20 %
Temperature Temperature of the sensor node, above the upper plot
Signal Strength RSSI converted into a distance relative to the maximum distance of the selected radio mode (linear, since v7.0)
Live sampling rates Measured live per channel — top: primary sensor / analog front end; bottom: IMU / digital inputs
Node overview Status, signal strength and sum sampling rate of all up to 8 nodes of the network
USB Buffer / Data Queue Data buffered while the software cannot keep up (e.g. high PC load). A rising value delays display and logging; if it does not fall again, use another PC

Mounting & Connections

nemi Connect physical connection

The nemi Connect is a USB device that appears on the PC as a virtual COM port (USB CDC ACM). No driver installation is typically required on Windows 10+.

  • Use a data-capable USB cable. Charge-only cables (common with bicycle lights and some accessories) have no data lines and will not establish a connection.
  • Avoid USB hubs if possible; connect directly to a PC USB port for most reliable enumeration.
  • The nemi Connect antenna should point vertically upward for best omnidirectional performance.
  • Keep the nemi Connect outside metal enclosures and metal cabinets. Metal strongly attenuates 2.4 GHz signals; even a thin metal wall can reduce effective range by 50% or more.
  • For installations where the sensor nodes are embedded in metal structures (nacelles, pylons, bridge boxes), position the nemi Connect near any opening in the structure, or at a location with a clear line of sight toward the sensors.

nemi EdgeBase network connection

The nemi EdgeBase connects to wirelessSensorPlot over TCP/IP on port 9413. The EdgeBase can be reached via:

  • Direct USB-Ethernet (for initial setup, no router needed):

PC IP address:    192.168.101.1 / 255.255.255.252
EdgeBase IP:      192.168.101.2
EdgeBase gateway: 192.168.101.1
Test with ping 192.168.101.2.

  • LAN — standard Ethernet switch/router connection.
  • LTE/VPN — the EdgeBase modem establishes a VPN; wirelessSensorPlot connects to the EdgeBase VPN IP. Ensure lte_check_ip: 8.8.8.8 is set in system_config.yml.

EdgeBase configuration files (YAML)

Access the EdgeBase configuration via WinSCP over FTP using the account nemibase_config. FileZilla is not recommended — WinSCP handles in-place edits and file transfers more reliably for EdgeBase YAML files.

File Purpose
system_config.yml Hostname, LAN IP, gateway, subnet mask, NTP servers, LTE PIN/APN/check IP
sensor_config.yml Internal sensors, radio network, logging mode, trigger settings, node parameters
mqtt_config.yml MQTT broker address, credentials, root topic, send interval

Detailed Configuration

Configuration layers

Layer Tool Stored where
Sensor node radio identity nemiSensorConfig (USB) On the sensor node (flash memory)
Runtime measurement settings wirelessSensorPlot Sent to sensor at receiver start; saved in *.ini on PC
Receiver / radio network wirelessSensorPlot Runtime; saved in *.ini on PC
EdgeBase autonomous behavior YAML files on EdgeBase EdgeBase filesystem

Radio network parameters

Parameter Valid values Where to set
Radio channel 0–79 (default 0) nemiSensorConfig (node), wirelessSensorPlot (receiver) — must match
Max nodes per channel 1, 3, or 8 nemiSensorConfig (node), wirelessSensorPlot — must match
Node number / ID 1–8 nemiSensorConfig (node) — must be unique per channel
Radio mode 2 / 1 / 0.5 / 0.125 Mbit/s nemiSensorConfig (node), wirelessSensorPlot — must match

Channel recommendation: Channel 0 is the default and a good choice. For parallel networks use channels 0, 25, 50, or 75 to minimize inter-network interference; numerically adjacent channels can bleed into each other. Channel 10 also works but may be slightly more exposed to Wi-Fi interference than channel 0.

Changing settings while running:

Change Method
Channel only Set Channel button — takes effect immediately
Radio mode or max nodes Disconnect COM-Port → wait → reconnect → change → Start Receiver
Node measurement parameters Change value in the tab, then click Update Node Settings (the button text turns red when there are unsent changes); the settings of all nodes are sent

A node only changes its configuration if the new settings differ from its current ones.

Automatic sensor-type correction: After start-up, all tabs are configured as nemi DAQ. If a tab is set to nemi DAQ but a nemi G+ transmits on this node number (or vice versa), wirelessSensorPlot switches the tab to the correct type as soon as the first data message arrives. During approx. the first 10 seconds a nemi G+ transmits at 4,000 Hz, and a nemi DAQ sends no analog data and may use a wrong IMU rate (caused by the radio protocol); after that, the stored settings are applied automatically.

8-node networks: The drop-down menu at the top left of each of the three tabs selects which of the 8 nodes is shown and configured there. The settings of the previously selected node are saved and restored automatically. If a node is selected on several tabs, its data is shown only on the rightmost of these tabs.

Node names (since v7.2): Click Set Sensor Names, enter a name per node and confirm with Set Sensor-Node Names. The names are shown in the software and used in the file names of recordings.

nemi G+ / nemi G+ ultra sensor parameters

Parameter Description
Range (primary MEMS) Measurement range of the primary accelerometer, shown as 2/10 g, 4/20 g, 8/40 g: the first value applies to the M08 variant (±2/4/8 g), the second to the M40 variant (±10/20/40 g)
Acc Range (IMU accelerometer) IMU accelerometer range: off, ±2 g, ±4 g, ±8 g, ±16 g
Gyr Range (IMU gyroscope) IMU gyroscope range: off, ±125, ±250, ±500, ±1,000, ±2,000, ±4,000 °/s
Mag Range (IMU magnetometer) Can be disabled (set to 0)
ODR Output data rate of the primary MEMS sensor
Mode 0 = Sleep, 1 = Continuous, 2 = Intermittent (timed modes: measure for Sample [s], then sleep for Sleep [s])

nemi G+ ultra additional modes:

Sample rate Max. active axes Max. nodes on channel
26.6 kHz 1 1
13.3 kHz 3 1
13.3 kHz 1 3
6.6 kHz 3 3
3.3 kHz 3 3

To disable the IMU entirely, set Acc Range, Gyr Range, and Mag Range all to 0. Keeping the IMU enabled is recommended for rotating applications — it provides orientation and angular velocity data without increasing radio bandwidth significantly.

For rotating machinery: verify the gyroscope range before deployment. At 250 rpm wheel speed with typical bearing geometry, roller angular velocity can reach 3,200–3,900 °/s; use ±4,000 °/s in such cases.

nemi DAQ analog input configuration

Unit Conversion performed Typical use
mV/V Normalized to 2.5 V reference Default; strain gauges (DMS)
V mV/V × 2.5 V Raw voltage
°C (Pt100) Automatic Pt100 resistance → temperature Pt100 thermometer (constant-current configuration)
°C (Pt1000) Same model; R₀ = 1,000 Ω Pt1000 thermometer (constant-current configuration)
0–10 V mV/V × 2.5 V × 8.647058824 DAQ channels wired for 0–10 V input

The 0–10 V unit requires a nemi DAQ configured for 0–10 V inputs (configuration option, see nemi DAQ); the attenuation of the 0–10 V input is taken into account in the conversion. Selecting the 0–10 V unit in software without the matching hardware configuration does not extend the input range. If different sensor types are connected to the channels, use V or mV/V as universal units and convert in post-processing.

DAQ gain settings

Gain Measurement range
1 ±1,000 mV/V
4 ±250 mV/V
8 ±125 mV/V
16 ±62.5 mV/V
32 ±31.25 mV/V
64 ±15.625 mV/V
128 ±7.8125 mV/V

Gain 1 is not recommended for routine use — it results in a low input impedance. For the recommended gain with Pt100 / Pt1000 sensors, see nemi DAQ. The displayed value in mV/V is always referred to the actual sensor signal regardless of gain — wirelessSensorPlot compensates the gain automatically.

Digital inputs (DIN) — nemi DAQ

DIN high-level setting Low threshold High threshold
3/5 V logic 0.8 V 2.0 V
12 V 3.5 V 8.5 V
24 V 7.0 V 17.0 V

All DIN channels are configured to the same threshold in wirelessSensorPlot. The DIN sampling rate is fixed at 2,000 Hz and is not user-adjustable.

Slope and offset scaling

wirelessSensorPlot applies a linear conversion y = a·x + b to every channel before display and logging. Set a (slope) and b (offset) per channel in the sensor settings panel. Changes take effect immediately. Values are saved in the *.ini settings file.

Saving and loading configurations

  • Save Settings to File — writes all current settings (radio channel, node parameters, unit selections, slope/offset, node names) to a plain-text *.ini file.
  • Load Settings from File — restores a saved configuration. If a loaded setting conflicts with the actual sensor type (detected from the first data packet), wirelessSensorPlot automatically corrects the mismatch within ~10 seconds.

The *.ini file can also be opened and edited with a text editor. If a loaded file specifies a different sensor type for a tab than the one actually connected, these settings are discarded as soon as data from that node arrives (see automatic sensor-type correction).

Settings are not stored on the sensor nodes (except radio parameters, which are flashed via nemiSensorConfig). Every session requires either loading an *.ini file or re-entering parameters manually.

Further functions: Pause Plotting freezes the live plots; Clear Plots and Consoles resets plots, consoles and further user-interface elements.

Intermittent mode (nemi DAQ)

For low-duty-cycle applications (e.g., quasi-static structural monitoring), configure the nemi DAQ with:

Setting Typical value
Mode Intermittent
Sample time Measurement window duration (e.g., 1 s)
Sleep time Rest duration (e.g., 4 s for ~1 Hz effective rate)
Sample rate Lowest High-Res rate applicable (e.g., 100 Hz)

"Samples missing" messages during intermittent mode are normal and do not indicate data loss; wirelessSensorPlot interpolates between burst windows.


Integration

Integration options

Path Best for Data type
nemi Connect + wirelessSensorPlot Local live display, CSV logging Live decoded data
EdgeBase + wirelessSensorPlot (TCP) Remote live display via VPN/LAN Live decoded data over TCP port 9413
EdgeBase local logging Long-term unattended measurements .bin / .bin.bz2
EdgeBase MQTT Dashboards, databases, aggregate monitoring Time-aggregated JSON messages
nemi Connect + Python API Custom applications, automated pipelines COBS-encoded binary
nemi Connect + Gantner Q.Station plugin Integration with GI Q.Station and GI.bench COBS-encoded binary

Gantner Instruments (GI) Q.Station

The nemi Connect can be connected directly to a GI Q.Station USB port and controlled via the GI Plugin for nemione (GINS.nemione). See the Gantner Knowledge Base for plugin documentation.

Recommended commissioning sequence:

  1. Configure all sensor nodes via nemiSensorConfig.
  2. Test the full setup in wirelessSensorPlot with the nemi Connect on a laptop — confirm stable reception, correct sample rates, no missing samples. wirelessSensorPlot's real-time Signal Strength indicator is much easier to use for RF diagnostics than GI.bench.
  3. Move the nemi Connect to the Q.Station USB port and re-enter the radio configuration in the GI Plugin.
  4. Run production data collection in GI.

USB protocol: The receiver is configured by the host (radio channel, radio mode, max. number of nodes, then one configuration message per node). The nemi Connect transmits raw bit codes; all unit conversions happen on the receiving end (wirelessSensorPlot, GI Plugin, nemi Studio or custom software). The measurement configuration (e.g. gain, range) is included in the data messages so the host can decode correctly. The complete protocol documentation is provided to customers on request and is disclosed to customers by i4M.

MQTT integration (nemi EdgeBase)

The nemi EdgeBase can publish measurement data to an MQTT or MQTTS broker.

mqtt_config.yml key settings:

Key Meaning
mqtt_use_mqtts 0 = plain MQTT, 1 = MQTTS (TLS)
mqtt_broker_address Broker address (URL or IP address)
mqtt_broker_port Port (commonly 1883 for MQTT, 8883 for MQTTS)
mqtt_broker_client_id Must be unique per client
mqtt_root_topic Root topic prefix
mqtt_send_interval Aggregation and send interval in seconds (e.g., 0.5 = 2 Hz)

Topics follow the pattern <mqtt_root_topic>/<sensor>/<stream> (e.g. …/nemi_DAQ_1/STREAM_0); <sensor> is the file_identifier of the sensor from sensor_config.yml, which is also used in the names of the logged files. Each message carries all channels of one stream as min / max / mean values over the send interval:

{
  "timestamp": "2026-06-24T19:43:52.077665+00:00",
  "values": {
    "CH_1": {"min_value": -0.97, "max_value": 0.12, "mean_value": -0.42}
  }
}
Sensor STREAM_0 STREAM_1 STREAM_2
nemi G+ / G+ nano / G+ ultra Vibration (VIB_X/Y/Z) IMU (ACC, GYR, MAG, TEMP, VOLT, RSSI) —
nemi DAQ / DAQ nano Analog channels (CH_1 …, VOLT, RSSI) IMU Digital inputs (DIN_1 …)
nemi EdgeBase Base IMU GPS / position GPIO / voltages (VOLTAGE_1 … VOLTAGE_4)

Every stream also contains TIMECOUNTER. See nemi EdgeBase for the complete MQTT documentation.

MQTT reliability considerations:

  • MQTT data is published once per interval as a live stream. If the network connection is down at publish time, that interval's data can be lost.
  • Use logging_mode: 4 (MQTT + local .bin.bz2) for any installation where data gaps are not acceptable. The raw files serve as a backup if MQTT delivery fails.
  • Two MQTT clients with the same client ID cause the broker to disconnect one. Use unique client IDs.
  • For MQTTS: certificates must be uploaded to the mqtts_certs/ folder via FTP (nemibase_config account) as ca.crt, client.crt, and client.key. The broker hostname in the configuration must match the certificate CN — using an IP address instead of the hostname causes TLS verification failure.

EdgeBase logging modes

logging_mode Output
0 Uncompressed .bin files
1 Compressed .bin.bz2 files
2 CSV files directly (risky at high data rates)
3 MQTT aggregates only — no local raw files
4 MQTT aggregates + .bin.bz2 raw archive
5 Triggered mode, direct CSV output
6 Live TCP stream on port 9413 to wirelessSensorPlot (v7.3+) or nemi Studio — no local files

Python API (nemi Connect)

i4M provides a Python API for direct access to the nemi Connect data stream, enabling custom pipelines and analysis workflows. It delivers the data of all nodes as pandas DataFrame queues; a graphical and a console example are included.

The API is provided free of charge without paid support. It is suitable for users who can build and maintain Python applications.

Time synchronization in wirelessSensorPlot

wirelessSensorPlot implements a two-stage synchronization model.

Stage 1 — Within the nemi Link 2400 wireless network: The clocks of all sensor nodes are continuously synchronized to the receiver (nemi Connect or nemi EdgeBase). The synchronization accuracy within the wireless network is ±10 µs with a stable radio connection. With a poor radio connection, some synchronization events have to be discarded because their transit time is too long, and the accuracy decreases.

In addition, wirelessSensorPlot relates the sensor time to the PC clock and stores the PC system timestamp for every sample. Establishing this relationship is what the 10-second wait before recording is for.

Stage 2 — From the receiver to the PC / target system: This is handled differently depending on the receiver:

  • With nemi Connect: The receiver is directly connected to the PC. PC-side timestamps written to CSV are derived from the Windows system clock (NTP-synced). Inter-PC sync quality depends on NTP, typically 1–50 ms — adequate for most applications but coarser than Stage 1. For high-precision cross-site comparisons, correct for NTP uncertainty in post-processing.
  • With nemi EdgeBase: The EdgeBase handles its own clock alignment to GNSS or NTP independently. This affects mainly multi-EdgeBase installations where globally aligned timestamps are required across sites.

Multi-receiver setups with wirelessSensorPlot:

  • Same PC: Connect all nemi Connect receivers to one measurement PC. All timestamps are derived from the same system clock — best achievable consistency.
  • Multiple PCs: Synchronize all PCs to the same NTP server before measurement. Residual inter-PC offset will be NTP-quality (~1–50 ms).

Downstream analysis tools

Tool Notes
FlowCSV Viewer Free (Microsoft Store); natively reads wirelessSensorPlot CSV format
PlotJuggler Open source; use Timecounter (s) as the X axis for import
Python (pandas, scipy, matplotlib) pd.read_csv(file, sep=';', decimal='.')
MATLAB / GNU Octave Standard readtable with delimiter settings
DIAdem (NI) Import via CSV with appropriate delimiter configuration

wirelessSensorPlot does not include a built-in FFT. A live FFT view is available in the successor software nemi Studio and the nemi Studio Web App.

Decoding EdgeBase binary files

  1. Download the .bin.bz2 files from the EdgeBase using WinSCP with the nemibase_data account. Files are organized in day-based folders.
  2. Decode them with the i4M Python decoder. i4M provides Python decoders uniformly for the efficient .bin.bz2 files of the nemi EdgeBase and the .bin files of the nemi Log. They decode the files into pandas DataFrames, which can be saved as CSV, Parquet etc. or processed directly in Python.

The former EdgeBase CSV decoder (.exe) is no longer maintained.


Troubleshooting

1. No COM port detected / nemi Connect not recognized

Symptom: wirelessSensorPlot or nemiSensorConfig shows no COM port after connecting the nemi Connect.

Causes and fixes:

  • Software was already open when the device was connected and the port was not detected — disconnect the device, wait for the software to fully start, then reconnect.
  • Another i4M tool has the COM port open — close all instances of wirelessSensorPlot and nemiSensorConfig, then reconnect.
  • USB cable does not support data — replace with a confirmed data-capable cable.
  • USB physical disconnect while the port was still open — close software, unplug, reopen software, reconnect.

2. No live data / sensor not found

Symptom: Plots remain empty after clicking Start Receiver; sensor LEDs blink rapidly or randomly.

Causes and fixes:

Cause Fix
Radio settings mismatch (channel, mode, or max-nodes) Verify all three parameters match between wirelessSensorPlot and sensor node (via nemiSensorConfig)
Two receivers active on the same channel Disable or reconfigure one receiver
Sensor off, sleeping, or discharged Power-cycle the sensor; check battery voltage
Receiver antenna inside a metal enclosure Move the nemi Connect outside the enclosure
Start Receiver not yet clicked Ensure the console shows "Waiting for Config Messages" before clicking Start Receiver
Duplicate node numbers on the same channel Assign unique node IDs via nemiSensorConfig

3. Missing samples (yellow or red status)

Symptom: Console reports "Samples missing"; background turns yellow or red.

Causes and fixes:

Cause Fix
RF link too weak for the selected mode Switch to a slower radio mode (500 kbit/s or 125 kbit/s); check Signal Strength
2.4 GHz interference (Wi-Fi, Bluetooth, smartwatches) Choose a less-congested channel; switch to a lower data rate
Sample rate too high for node count + radio mode Reduce sample rate or number of active channels
Antenna orientation or obstruction Point nemi Connect antenna vertically; remove from metal enclosures
Low battery Check battery voltage; LiPo 4.1 V = ~100%, 3.3 V = ~20%
PC overloaded Watch the USB Buffer / Data Queue counter; if it grows continuously, use a faster PC

4. Wrong measurement values

Symptom: Values are at the wrong magnitude, sign, or show a constant offset.

Causes and fixes:

  • Range setting not applied: After changing a range, click Update Node Settings (button turns red when changes are pending). New settings take effect ~10 seconds later.
  • Wrong primary sensor range: Verify that the range selected in the UI covers the physical signal; a too-narrow range can appear as clean data at an unexpected scale without obvious saturation.
  • Open DAQ input (no sensor connected): An unconnected differential input drifts arbitrarily. Connect a sensor or terminate the input before evaluating measurements.
  • Gyroscope range saturated in rotating application: Use ±4,000 °/s for high-speed applications; ±2,000 °/s clips at ~500 rpm depending on the component geometry.

5. Spikes in nemi DAQ data

Symptom: Short voltage spikes appear in analog data, particularly with DMS (strain gauges) at low sample rates (≤ 100 Hz).

Cause: In Low Power mode (≤ 100 Hz), the DMS bridge is powered in short pulses. If the bridge cable has capacitance, the bridge voltage may not fully settle before sampling.

Fix: Switch to a High-Res mode (≥ 100 Hz continuous), where the bridge power is permanently on. Spikes disappear. With 1,000 Ω DMS bridges, the power consumption difference is modest.

6. 1 Hz sampling not working as expected (nemi DAQ)

Symptom: At 1 Hz, data arrives only every ~66 seconds (single channel) or not at all.

Cause: The nemi DAQ firmware fills packets to a fixed sample count before transmitting. At 1 Hz with one channel active, filling a packet takes ~66 seconds.

Fix (preferred): Use Intermittent mode with a higher sample rate:

Mode: Intermittent
Sample time: 1 s
Sleep time: desired interval (e.g., 4 s)
Sample rate: 100 Hz (or lowest High-Res rate)
Alternatively: enable all three analog channels — even with only one carrying a signal — to reduce the packet-fill time at low rates.

7. Intermittent connectivity / EdgeBase VPN drops

Symptom: The EdgeBase appears in the VPN network, then drops out. Ping intermittently fails.

Cause: A misconfigured lte_check_ip in system_config.yml causes the EdgeBase to incorrectly conclude its LTE connection has failed and restart it.

Fix: Set lte_check_ip: 8.8.8.8 in the EdgeBase configuration. Access via FTP (nemibase_config account) and edit the YAML.

8. EdgeBase live connection — no local files created

Symptom: wirelessSensorPlot connects to the EdgeBase but no files appear on the EdgeBase.

Cause: logging_mode: 6 is active. This mode streams live data to wirelessSensorPlot and does not create local raw files.

Fix: For local file logging: use logging_mode: 1 (binary archive) or logging_mode: 4 (binary archive + MQTT). Use logging_mode: 6 only for live wirelessSensorPlot sessions.

9. MQTT data gaps

Symptom: MQTT messages are missing or gaps appear in downstream dashboards; raw EdgeBase files appear complete.

Causes and fixes:

  • MQTT data is sent once per interval and not re-queued on failure. Use logging_mode: 4 so .bin.bz2 files are also stored; raw data remains available for gap-filling.
  • Reduce the MQTT send interval to reduce data loss per connectivity event; compute averages server-side in the dashboard.
  • For MQTTS: verify certificate paths and that the broker hostname matches the certificate CN (not the IP address).
  • Two MQTT clients sharing the same client ID cause one to be disconnected; use unique client IDs.

10. Battery protection fuse tripped (nemi G+)

Symptom: A nemi G+ no longer powers on after a USB cable issue (possible short circuit).

Cause: The integrated LiPo battery has an internal protection fuse that trips under a short circuit.

Fix (hardware):

  1. Open the housing (Torx screws).
  2. Remove the inner screws and lift the top PCB.
  3. Momentarily shift the top PCB one pin outward in any direction, then re-seat it correctly.
  4. Test with housing open; if recovered, reassemble.

11. Trigger threshold missed in EdgeBase trigger mode

Symptom: In trigger-based EdgeBase logging, the trigger stops firing even though the physical event continues.

Causes and fixes:

  • Temperature-induced baseline drift: The IMU in sensor nodes has measurable temperature dependence. A seasonal or day/night temperature swing can shift the resting signal by enough to push it below the configured threshold. Adjust trigger_threshold in sensor_config.yml with a temperature margin.
  • Depleted trigger sensor battery: Check for increasing sample loss on the trigger sensor node; if confirmed, switch to a different node and update trigger_sid in the EdgeBase YAML.

Systematic troubleshooting approach

  1. Start with a close-range test (< 1 m) using wirelessSensorPlot + nemi Connect. If data arrives cleanly, the radio configuration is correct and the problem is range/RF-related.
  2. Use the Signal Strength indicator to quantify reception quality. 0 % means the connection is at its limit; even receiving data at 0 % is marginal.
  3. Save debug_x.csv files for all affected sensor slots and send them to i4M if the issue cannot be resolved. They contain full console output including firmware error codes.
  4. When investigating data quality, isolate the issue to the live view (eliminates post-processing as the source); if only post-processed data shows artifacts, the problem is in decoding or MQTT delivery, not in the sensor system.

Product Changes

wirelessSensorPlot version history

Version Release Key changes
7.0 2023 New calculation of the Signal Strength indicator (RSSI converted into a distance relative to the maximum distance of the radio mode, linear)
7.1 Beta 4 January 2024 Bug fixes for settings load/save; auto sensor-type correction — if a slot is configured as one sensor type but a different type connects, wirelessSensorPlot auto-corrects within ~10 s and selects a compatible sample rate automatically
7.2 2024 Node naming — assign custom names per node; names appear in the UI and in CSV filenames
7.3 October 2024 EdgeBase live TCP stream — wirelessSensorPlot can connect to a nemi EdgeBase over LAN or VPN via TCP port 9413; EdgeBase must be set to logging_mode: 6; EdgeBase IMU available as a data source
7.4 Beta 3 January 2026 (last version) Full support for nemi G+ ultra and nemi G+ ultra nano, including the new USB data message format; backward-compatible with existing nemi G+ and nemi DAQ setups
— 2026 wirelessSensorPlot is no longer maintained; replaced by nemi Studio and the nemi Studio Web App

USB protocol documentation

The USB data messages of all nemione® USB devices (nemi Connect, nemi Log, nemi EdgeBase) are described in the i4M protocol documentation Dokumentation USB Datennachrichten, current version 2026-05-04. It is relevant for third-party integration developers and is provided to customers on request.

Sensor node firmware updates

Sensor node firmware (nemi G+, nemi G+ ultra, nemi DAQ, nemi DAQ nano) is updated by i4M via USB. Options:

  • Ship the device to i4M.
  • On-site visit from i4M.
  • Remote-guided update if USB access is available at the installation site.

Firmware updates do not change radio channel settings; the node retains its configured channel, node number, and radio mode after an update.

EdgeBase software updates

EdgeBase software is deployed remotely by i4M via VPN. The customer connects the EdgeBase to a VPN accessible by i4M; i4M logs in and deploys the update. Features such as logging_mode: 6 (TCP live stream) require an EdgeBase software update for units shipped before October 2024.

Compatibility notes

Scenario Note
Old nemi G+ + new wirelessSensorPlot Fully compatible; wirelessSensorPlot detects firmware version from first data packet and sends the appropriate configuration message format
Old nemi G+ + new protocol message sent by mistake Sending a new-format range/axis configuration to an old sensor sets an unintended range; wirelessSensorPlot handles this automatically
wiredSensorConfig (predecessor to nemiSensorConfig) Deprecated; do not use for new setups
*.ini settings files across minor versions Compatible across minor wirelessSensorPlot versions

FAQ

Applications

Q: Is wirelessSensorPlot suitable for inclinometry / tilt measurement?

The nemi G+ can be used as an inclinometer with post-processing (moving median filter, long averaging window). With calibration: resolution ~0.001° (17.45 µm/m), accuracy ~0.05° over the full 0–360° range. The sensor is well-suited for monitoring changes in inclination (structural monitoring); for absolute angle, long-term thermal drift must be accounted for. Use High-Res sample rate modes and a moving-median window > 1 second.

Q: Can I use wirelessSensorPlot for bearing and rotating machinery monitoring?

Yes. Verify the gyroscope range before final tests: at 250 rpm, bearing roller angular velocity can reach 3,200–3,900 °/s — use ±4,000 °/s to avoid saturation. The nemi G+ nano has been used in rotating components (rotor blades, bearing rollers) without modification.

Q: Can I run long-term unattended measurements?

wirelessSensorPlot requires the PC to remain connected and running. For unattended long-term measurements, use the nemi EdgeBase with logging_mode: 1 or logging_mode: 4. wirelessSensorPlot can be used to commission the setup and for spot checks during the campaign.

Q: Can wirelessSensorPlot be used without a nemi Connect?

Yes, if you connect to a nemi EdgeBase configured with logging_mode: 6 and reachable over the network. For direct live radio reception on a PC, use a nemi Connect.

Connectable sensors

Q: Can I mix nemi G+ and nemi DAQ nodes on the same radio channel?

Yes. Up to 8 nodes per channel are possible; each node can be a different type. The software detects the sensor type from the first data packet and applies appropriate settings automatically (after approx. 10 seconds). For mixed networks, configure each node individually and save the settings to an *.ini file.

Q: Can I connect Pt100 and strain gauges (DMS) to the same nemi DAQ nano?

No. Strain gauges use constant-voltage excitation (2.5 V); Pt100/Pt1000 use constant-current excitation. These are hardware-configured modes that cannot be mixed on one DAQ nano. Use one DAQ nano per sensor family. Exception: thermistors (NTC/PTC) can share a DMS-configured DAQ nano since they are passive resistive sensors.

Q: Should I choose the 8 g or the 40 g nemi G+ variant?

The 8 g variant (M08) has a minimum range of ±2 g — lowest noise, best resolution for small signals. The 40 g variant (M40) has a minimum range of ±10 g — suitable for machinery, shock, and hammer excitation. Start with the 8 g variant for civil and structural applications; request the 40 g variant only when physical accelerations clearly exceed ~6–8 g.

Q: Can I mount the nemi G+ magnetically?

The datasheet allows screw, adhesive, or magnetic mounting. A grub-screw connection provides the best vibration transmission. Magnetic mounting affects the built-in magnetometer reading.

Power supply

Q: What USB cable do I need for the nemi Connect?

Use a data-capable USB cable. Charge-only cables — common with bicycle lights and some power banks — have no data lines and will not work.

Q: The nemi DAQ nano restarts repeatedly. What causes this?

The most common cause is an unstable or insufficient supply. Power supplied over Micro USB is less reliable than direct supply-pin wiring, especially for permanent installations. Use the specified supply pins and voltage range. Also check that node numbers are unique on the channel.

Q: How long does the battery last?

Battery life depends on radio mode, sample rate, and number of active channels. For general guidance: LiPo battery at ~4.1 V is ~100% charged; at ~3.3 V it is ~20%. Increasing sample rate or switching to a faster radio mode increases consumption. For exact figures, contact i4M with your specific configuration.

Range

Q: How far can sensor nodes transmit?

See the radio mode range table in Technical Specifications and nemi Link 2400. For 125 kbit/s: ~350 m in urban environments, ~140 m in buildings. Metal structures significantly reduce these figures; position the receiver near structural openings.

Q: Which radio mode should I use for rotor blades or nacelles?

2 Mbit/s is suitable only for short distances (~20 m). For structures where RF conditions are challenging, start with 500 kbit/s or 125 kbit/s. Choose the slowest mode that still provides the required sample rate.

Q: Can two nemi Connect receivers listen on the same channel simultaneously?

No. Two active receivers on the same channel cause RF collisions and data loss. Assign different channels to different sensor groups, or use a single receiver with 8-node mode.

Integration

Q: How many sensors can I run simultaneously?

Up to 8 sensor nodes per radio channel. wirelessSensorPlot displays three nodes at a time across three tabs; all 8 are active simultaneously. For more than 8 sensors, use multiple channels — each requiring a separate receiver (nemi Connect or EdgeBase).

Q: Why does wirelessSensorPlot show only three tabs if the network can contain eight nodes?

The three tabs are display windows. In 8-node mode, each tab can be switched to any of the eight nodes; settings for previously displayed nodes are saved automatically when you switch. If a node is selected on several tabs, its data is shown only on the rightmost one.

Q: Does MQTT contain full-resolution data?

No. EdgeBase MQTT publishes time-aggregated values (min/max/mean) at the configured send interval. For full-resolution data, use EdgeBase binary logging (logging_mode: 1 or logging_mode: 4) and decode with the i4M Python decoder (pandas DataFrames → CSV, Parquet etc.).

Q: Can I compare wirelessSensorPlot CSV output against data from a third-party plugin?

Yes. wirelessSensorPlot is the reference tool. If wirelessSensorPlot logs correctly but a plugin does not, compare channel mode, node count, radio speed, chunking behavior, and data-format assumptions. The CSV timestamp columns and the i4M USB protocol documentation (provided to customers on request) are the common reference.

Q: Which timestamp column should I use for analysis?

Use Timecounter (s) for most single-recording analyses — it is not affected by PC clock corrections during the recording. Use Timestamp (s) when aligning data to an absolute UNIX time base.

Q: Which CSV import settings should I use?

Semicolon (;) as separator and period (.) as decimal separator.

Q: How can I view recorded data quickly?

The free FlowCSV Viewer (Microsoft Store) opens the CSV files directly. For further analysis use e.g. GNU Octave, MATLAB or Python.

Q: Can several nemi Connect receivers record time-synchronously?

Yes. Connect them to the same measuring PC — every sample is stored with the PC system timestamp. Alternatively, use several PCs synchronized to the same NTP time server.

Q: My settings are not saved after restarting wirelessSensorPlot.

Settings are not stored on the sensor node (only the radio parameters are, via nemiSensorConfig). Use Save Settings to File to write a *.ini file and Load Settings from File to restore it at the next session.