nemi Connect¶

Description¶
nemi Connect is a compact wireless receiver that bridges up to eight nemi Link 2400 sensor nodes to a PC or other host device via a single USB-C connection. It acts as the hub of a star-topology wireless measurement network: sensor nodes transmit acceleration, strain, temperature or digital signals over the 2.4 GHz nemi Link 2400 radio protocol; nemi Connect receives those packets and streams the measurement data directly to the host as a standard USB serial device.
The device is powered entirely by USB-C (5 V) and appears on the host as a USB-CDC virtual COM port — no proprietary driver is required on Windows 10/11 and other operating systems. The included software nemi Studio (Windows) and the browser-based nemi Studio Web App provide receiver and sensor configuration, live data display, live analysis and CSV recording. A Python API is available for integration into any data pipeline.
Within the nemione product family, nemi Connect is the PC-tethered receiver, suited to measurements where a laptop, desktop PC or other USB-ready host device such as a Gantner Instruments Q.Station is present at or near the measurement site. For autonomous long-term logging without a permanent PC connection, the nemi Log or nemi EdgeBase receivers are the appropriate choice. A nemi Log can also be used as a USB receiver in the same way as a nemi Connect, e.g. with nemi Studio or the nemi Studio Web App.

Typical Applications¶
- Test bench and laboratory measurements — nemi Connect plugs directly into the PC running the measurement software; no additional hardware is needed.
- Prototype and development setups — rapid configuration, live signal inspection and live analysis (FFT, attitude, rotation angle) with nemi Studio or the nemi Studio Web App.
- Field commissioning and radio range tests — before finalizing an installation with nemi Log or nemi EdgeBase, use nemi Connect and a laptop to validate reception quality, sampling rates and signal integrity at the actual site.
- Wired DAQ and host system integration — when wireless nemione sensor data must be combined with wired channels, connect nemi Connect to a Gantner Instruments Q.Station (GI.bench, GI.cloud, Gantner fieldbus output). nemi Connect is also the interface for integrating high-performance wireless sensor networks into your own host device.
- Custom software integrations — the documented USB binary protocol and the provided Python API enable integration into Python pipelines, SCADA bridges or OPC-UA gateways.
Technical Specifications¶
General¶
| Parameter | Value | Unit |
|---|---|---|
| Dimensions (without antenna and connectors) | 58 × 45 × 28 | mm |
| Weight | approx. 60 | g |
| External power supply | 5 (USB-C) | V |
| Data connection to the PC | USB-C to standard USB 2.0 | — |
| Antenna connector | RP-SMA, 2.4 GHz antenna included | — |
| Software for settings, live display and data storage | nemi Studio, nemi Studio Web App & Python API, included | — |
| Operating temperature | −20 to 80 | °C |
| Housing protection class | IP 41 | — |

Radio (nemi Link 2400)¶
| Parameter | Value |
|---|---|
| Frequency band | 2.4 GHz ISM |
| Network topology | Star (one receiver, up to 8 nodes) |
| Supported radio modes | 2 Mbit/s, 1 Mbit/s, 500 kbit/s, 125 kbit/s |
| Max. sensor nodes per receiver | 8 |
| Max. range | up to 350 m (depending on radio mode and environment) |
| Clock synchronization accuracy (sensor nodes to receiver) | ±10 µs with a stable radio connection |
| Latency (node to USB host) | typ. < 20 ms |
Compatible Sensor Nodes¶
nemi Connect is compatible with all sensor nodes in the nemi Link 2400 network:
| Sensor nodes | Sensor nodes (OEM / PCB versions) |
|---|---|
| nemi G+ | nemi G+ nano |
| nemi G+ ultra | nemi G+ ultra nano |
| nemi DAQ | nemi DAQ nano |
| nemi Struct | nemi Motion |
First Commissioning¶
Required Hardware and Software¶
| Item | Purpose |
|---|---|
| nemi Connect receiver | Receiver; USB-C cable included |
| Sensor nodes | e.g. nemi G+, nemi DAQ or any other nemi Link 2400 node |
| Host PC / device | Windows 10 or higher for nemi Studio; any OS with Chrome or Edge for nemi Studio Web App; any OS with the Python API; USB 2.0 or higher |
| nemi Sensor Config v2.1 or later | Persistent radio parameters of the sensor nodes via USB |
| nemi Studio v1.1 or later | Windows desktop software: receiver and node setup, live display, live analysis, CSV recording |
| nemi Studio Web App | Browser alternative: open https://nemi.one/studio/ in Chrome or Edge (Web Serial), no installation |
nemi Studio and nemi Sensor Config require no installation: unzip the package into a folder with write access (e.g. below Documents, not into C:\Program Files) and run the .exe. Download links are provided to customers; the software is not publicly available.
Step 1 — Plan the Radio Network¶
Before handling hardware, decide:
| Parameter | Typical values |
|---|---|
| Radio channel | 0, 25, 50 or 75 (see Radio Channel Selection) |
| Radio mode | 2 Mbit/s, 1 Mbit/s, 500 kbit/s or 125 kbit/s |
| Max. node count | 1 (low latency), 3 or 8 sensor nodes |
| Node numbers | 1 – 8, unique per sensor |
All sensor nodes and the receiver must share the same channel, radio mode and max. node count. Use the smallest network mode that covers the actual number of nodes — e.g. do not use 8-node mode for two nodes when the sampling-rate budget is tight.
Step 2 — Configure Each Sensor Node via USB¶
Connect each sensor node individually to the PC and open nemi Sensor Config. Set:
- Radio channel (same on all nodes)
- Radio mode (same on all nodes)
- Max. node count (same on all nodes)
- Node number (unique, 1 – 8)
See the commissioning section of the respective sensor node (e.g. nemi G+) for the detailed procedure.
Step 3 — Connect the Receiver with nemi Studio¶
- Disconnect the nemi Connect from the PC and start nemi Studio.
- Plug the nemi Connect into a USB port. The LED of the nemi Connect flashes red slowly. nemi Studio detects the new receiver and automatically selects its COM port.
- Click Connect COM-Port and wait until the message "Waiting for Config Messages to start …" appears in the Debug Console.
- Optional: load previously saved settings with Load Settings from File.
- Select the radio data rate and the number of sensor nodes per channel. These must match the settings written to the sensor nodes with nemi Sensor Config. Set the radio channel.
- Select the module type and the settings (sampling rate, ranges, mode) of each sensor node.
- Click Start Receiver. The receiver is configured and the settings are sent to all sensor nodes.
- Switch on the sensor nodes. Live data appears in the plots and the nemi Connect flashes green.
- Wait approx. 10 s for the initial PC ↔ node time synchronization; Start Recording is enabled afterwards.
Alternative — Connect with the nemi Studio Web App¶
- Plug the nemi Connect into the computer via USB.
- Open
https://nemi.one/studio/in Chrome or Edge (Web Serial API required; Safari and older Firefox versions cannot connect to a nemi Connect). - Configure the radio network and the sensors in the Node / radio settings area.
- Press Connect and select the port of the nemi Connect in the browser dialog. The app restarts the nemi Connect, sends the settings to the receiver and the sensors (approx. 10 s) and then waits for sensor data.
- The sensors typically need 10 – 20 s to join. As soon as data arrives, the status line turns green and a column appears for each sensor.
The Web App runs entirely in the browser; measurement data and recordings never leave your device.
Step 4 — Verify Data Acquisition¶
All expected nodes should appear with live data. Check the signal strength indicator — values below approx. 20 % indicate a marginal link. Check the node status indicators in nemi Studio:
| nemi Studio node status | Meaning |
|---|---|
| Green | All data streams complete |
| Yellow | Last sample older than 2 s (node may deliver buffered data later, or is in sleep mode) |
| Red | At least one sample lost within the last second — check reception or reduce the sampling rate |
Start a short CSV recording and check the files before mounting the sensors at their final location.
LED Status Codes¶
| LED behaviour | Meaning |
|---|---|
| Red flashing (slow) | Receiver powered, software not connected |
| Rapid green flashing | Data transmission ongoing |
| Off | No power |
Radio Channel Selection¶
Recommended channels: 0, 25, 50, 75
Channel 0 is the default and a good choice for most installations. Other channels work as well; some of them (e.g. channel 10) may be slightly more affected by Wi-Fi traffic than channel 0. When running multiple independent nemi Link 2400 networks simultaneously, assign each a different channel from this list. If you want to run more than 4 networks, keep at least 10 channels between the networks and avoid a spacing of 8 channels. A quick Wi-Fi scan at the site (e.g. a smartphone Wi-Fi analyser app) helps to assess the local 2.4 GHz environment.
Antenna Placement¶
- Keep the nemi Connect antenna upright wherever possible.
- Avoid placing the receiver or sensor antennas inside metal enclosures — a steel housing acts as a Faraday cage and can block 2.4 GHz radio entirely.
- If the receiver must be inside a metallic structure, route the antenna through a small opening. Avoid long RF cable extensions — they reduce the link budget significantly. Extending the USB cable is the preferred approach for difficult installations.
- Line of sight between receiver and sensors improves reliability. Concrete walls, metal structures and wet materials attenuate the signal.
- Do not hold sensors in your hand during range tests; the human body strongly absorbs 2.4 GHz radio.
- For outdoor or long-range deployments, switch to 500 kbit/s or 125 kbit/s to increase receiver sensitivity (at the cost of data throughput).
Mounting & Connections¶
nemi Connect has no fixed mounting interface; it is a desktop-placed unit. The physical connections are:
- USB-C port — power and data (USB 2.0). The supplied USB-C to USB-A cable plugs into a standard PC USB-A port.
- Antenna connector — 2.4 GHz antenna with RP-SMA connector.
Connect nemi Connect directly to a USB port of the host. USB hubs with integrated Ethernet adapters have caused intermittent disconnects; if a hub is unavoidable, use a powered hub without Ethernet functionality and test it before deployment.
For wet or harsh environments, house nemi Connect in a sealed enclosure (the device itself is IP41). A 5 m USB extension cable from the sealed enclosure to the host controller has been used successfully in difficult installations; verify USB cable length compatibility with the target controller before finalizing cable routing. If you can choose between a USB extension cable and an antenna extension cable, always extend on the digital USB side, as this optimizes radio reception.
There are no external sensor connections on nemi Connect itself. All sensor signals travel wirelessly; sensor nodes connect to their transducers (strain gauge bridges, RTDs, etc.) via their own wiring.
Detailed Configuration¶
Configuration is split into two layers:
- Persistent radio settings of the sensor nodes — written via nemi Sensor Config over USB (channel, radio mode, node number, max. node count).
- Receiver and runtime measurement settings — set in nemi Studio, the nemi Studio Web App or via the Python API (radio channel, radio mode and node count of the receiver; sampling rates, ranges, operating mode of each node). The settings are sent wirelessly to the nodes.
Radio Modes¶
| Radio mode | Typical use case | Sum sampling rate (24-bit samples) | Approx. range (indoor / urban) |
|---|---|---|---|
| 2 Mbit/s | Short range, maximum throughput, high sampling rates | 36 kHz | ~20 m / ~50 m |
| 1 Mbit/s | Balanced default for multiple nodes at moderate range | 15 kHz | ~50 m / ~125 m |
| 500 kbit/s | Extended range | 8 kHz | ~90 m / ~225 m |
| 125 kbit/s | Maximum range, lowest data rates | 3 kHz | ~140 m / ~350 m |
The achievable sampling rate depends on the radio mode, the network mode, the number of active nodes and channels, the message type, the radio quality (retransmissions) and the processing capacity of the host. For structural monitoring with dominant frequencies around 50–80 Hz, sampling rates of 250–500 Hz are common. High-rate sensors such as nemi G+ ultra may require single-node operation or fewer active axes at the highest sampling rates.
These are indicative figures. Metal structures, walls and wet materials reduce range; the reliable range must always be validated on site. Use the nemi Link 2400 Radio Channel Calculator to verify that the planned combination of sensors, sampling rates and radio mode fits within the channel bandwidth. See nemi Link 2400 for details.
Changing Radio Settings During Operation (nemi Studio)¶
- Radio channel: after Start Receiver, the channel can be changed at any time with Change Channel; the change takes effect immediately.
- Data rate and number of nodes: click Disconnect COM-Port, wait until the nemi Connect flashes red again, click Connect COM-Port, change the settings and click Start Receiver.
Node Operating Modes¶
| Mode | Description |
|---|---|
| 0 = Sleep | Node sleeps and wakes periodically to check for new settings; LED blinks green every ~10 s |
| 1 = Continuous | Node samples and transmits continuously |
| 2 = Intermittent | Node measures for a configured time, then sleeps for a configured time |
Settings changes are sent with Update Node Settings (nemi Studio) or Send settings (Web App). A node only applies settings that differ from its current ones; it may take up to approx. one minute until a node applies new settings.
Sensor Ranges and Scaling¶
nemi Connect has no measurement range of its own; range, resolution and excitation are defined by the connected sensor nodes. Choose the smallest range that does not clip. For low-amplitude vibration or inclination, ±2 g usually gives the best resolution — remember that gravity already uses about 1 g on the vertical axis. Magnetometer-based heading (yaw) can be disturbed by nearby steel or magnetic structures. For nemi DAQ nodes, check the hardware input configuration before connecting analog signals (see nemi DAQ).
Saving Settings¶
nemi Studio stores all settings (radio settings, node settings, node names, zero/scaling values, recorded channels) in an INI file via Save Settings to File. Save the configuration immediately after a verified commissioning and share it with team members to ensure consistent settings when resuming a measurement campaign.
Integration¶
nemi Connect exposes sensor data through several downstream paths. The primary interface is the USB virtual COM port; all other integrations build on top of it.
Time Synchronization¶
nemi Connect participates in a two-stage synchronization model.
Stage 1 — Within the wireless sensor network: All sensor nodes synchronize their internal clocks to the nemi Connect receiver via the nemi Link 2400 radio protocol, with an accuracy of ±10 µs with a stable radio connection. With a poor radio connection, some synchronization events have to be discarded because their transit time is too long, and the accuracy decreases. This is the primary high-precision timestamp source for multi-sensor correlation. The synchronized timestamp is embedded in every measurement data message.
Stage 2 — From nemi Connect to the host system: nemi Studio and the nemi Studio Web App relate the receiver time to the clock of the computer and assign a synchronized timestamp to each sample. In a Gantner Q.Station setup, the data is aligned to the GI DAQ network clock. Synchronize the computer clock with a time server (NTP). Several receivers on one computer, or several computers synchronized to the same NTP server, can thus record time-synchronously.
For this, the receiver regularly sends timing messages with its current radio network time. Custom integrations use them to map the sensor timestamps to the clock of the host (wall-clock time); the i4M Python API does this automatically.
USB COM Port Interface¶
After connecting the USB-C cable, nemi Connect appears as a USB CDC ACM virtual COM port (Device Manager → Ports (COM & LPT) → USB Serial Device; no proprietary driver required on Windows 10/11 or Linux).
The data is transmitted in a binary protocol that is shared by all nemione® USB devices:
- After the COM port has been opened, the receiver immediately sends info messages (human-readable debug output) and timing messages, and then waits for configuration ("Waiting for Config Messages to start …").
- The host first configures the receiver (radio channel, radio data rate, max. number of nodes) and then sends one configuration message per sensor node of the network (1, 3 or 8, depending on the network mode — also for nodes that are not active). A short pause between the configuration messages is recommended.
- The receiver then forwards the data messages of all sensor nodes, together with info and timing messages.
- All messages are COBS-encoded (Consistent Overhead Byte Stuffing), so that a zero byte only appears as the delimiter between messages.
- Each data message carries the synchronized timestamp (1 µs resolution) of the last sample in the message; the timestamps of the earlier samples follow from the sampling rate.
The complete protocol documentation (message formats and conversion to physical units) is provided to customers on request and is disclosed to customers by i4M.
Timer Counter Overflow¶
The radio network time is a 32-bit counter at 1 MHz (1 µs resolution). It overflows after 2³² µs ≈ 71.58 minutes (1:11:34.97 h).
In normal continuous operation, nemi Studio and the i4M Python API handle this transparently. When sensor data was buffered in the sensor's internal memory (sensor out of radio range) and later replayed, replayed timestamps may jump by exactly ≈ 4,294.967 s (≈ 71.6 minutes) at an overflow boundary. These discontinuities can be identified algorithmically and corrected in post-processing.
Python API¶
The nemi Connect Python API (provided by i4M on request) handles COM port access, COBS decoding and message parsing, and exposes the sensor data as pandas DataFrame queues in real time.
Getting started:
- Connect the nemi Connect to the PC via USB.
- Download the Python API package (ask i4M for a link).
- Open the included example script as a working reference.
The package includes example scripts and the sensor channel definitions; the protocol documentation is provided together with it on request. The Python API runs on Windows, Linux and macOS wherever USB-CDC is supported — any system with a USB port (e.g. Raspberry Pi or industrial PC) can host a custom integration. Implementation support for custom integrations is available on a per-project basis.
Gantner Instruments (GI.bench) Integration¶

With nemi Connect, wirelessly transmitted data can be displayed directly in GI.bench and GI.cloud. nemi Connect is connected to a Gantner Instruments controller via USB cable:
The nemione plugin makes nemi Connect channels available as virtual channels within GI.bench, from which data can be routed to Gantner fieldbus modules (Profibus, Profinet, EtherNet/IP, etc.), cloud storage or visualization tools.
Key advantages:
- Digital data remain digital end-to-end.
- Direct time synchronization via the GI DAQ network.
- Data feed directly into the GI DAQ network / cloud.
- Fieldbus output handled on the Gantner side.
Setup:
- Configure sensor nodes with nemi Sensor Config.
- Validate the radio network with nemi Connect and nemi Studio on a laptop.
- Confirm stable reception and correct sampling rates.
- Connect nemi Connect to the Q.Station via USB. Several nemi Connect units per Q.Station are possible.
- Install the nemione plugin and configure channel, radio mode and node settings.
- Start the Gantner project and compare the first signals against the nemi Studio reference.
Plugin configuration is independent of the nemi Studio configuration — set channel and radio mode explicitly in the plugin. If the Q.Station project becomes overloaded, reduce sampling rates temporarily and compare the behaviour against nemi Studio to isolate plugin/controller issues from radio issues.
PLC / Fieldbus Integration¶
nemi Connect has no native fieldbus interface. Two practical paths:
- Via Gantner Q.Station — nemi Connect provides data to the Q.Station via USB; the Q.Station routes selected channels to a PLC via its fieldbus module. This is the recommended path when real-time fieldbus output is required.
- Via Python bridge — use the Python API to read sensor data, then forward it to a PLC using an appropriate library (e.g.
python-snap7for Siemens S7, or an OPC-UA server library).
CAN is generally not recommended for high-rate raw sensor data as its bandwidth is insufficient.
Multi-Receiver Setups¶
When more than 8 sensor nodes are required, or when sensor groups need different radio modes (e.g. a high-speed group at 2 Mbit/s and a long-range group at 500 kbit/s):
- Deploy multiple nemi Connect units, each on a distinct radio channel (e.g. 0, 25, 50, 75).
- Connect each to a separate USB port (or separate PC).
- Use the Python API to read from multiple COM ports simultaneously, or run one nemi Studio instance per receiver.
- Synchronize all PCs to the same NTP server for aligned timestamps.
Data Pipeline Summary¶
| Use case | Recommended path |
|---|---|
| PC-based live measurement, live analysis and CSV recording | nemi Connect + nemi Studio |
| Quick live check / recording without software installation | nemi Connect + nemi Studio Web App (Chrome / Edge) |
| Custom software / automation | nemi Connect + Python API |
| Gantner / GI.bench integration | nemi Connect + Q.Station + GI nemione plugin |
| PLC / fieldbus (real-time) | nemi Connect + Q.Station + fieldbus module |
| PLC / fieldbus (custom) | nemi Connect + Python API + PLC library |
| IoT / remote monitoring, MQTT | nemi EdgeBase |
| Standalone logging without PC | nemi Log or nemi EdgeBase |
CSV Output (nemi Studio)¶
nemi Studio writes semicolon-separated CSV files; the first line contains the column names. Each sample has three time columns: Timestamp (ASCII) (local time, µs resolution), Timestamp (s) (UNIX time) and Timecounter (s) (time since start of recording). Files are written per node (e.g. <datetime>_<node>_vibration_data.csv, <datetime>_<node>_imu_data.csv, <datetime>_<node>_telemetry_data.csv); optionally a resampled common file of all nodes is created. See nemi Studio for the full file format.
Excel note: On European locales, use Data → From Text/CSV and select semicolon as delimiter rather than double-clicking the file. For a quick look at large files, use a dedicated CSV viewer or Python (pandas), MATLAB / GNU Octave.
Data Volume Estimates¶
At 300 Hz sampling rate per sensor (COBS-encoded binary USB stream):
| Data type | Approx. bytes/s per sensor |
|---|---|
| Analog / strain gauge channels (nemi DAQ, 3 channels) | ≈ 3,460 |
| IMU channels | ≈ 4,760 |
Example — 8 sensors, strain gauge + IMU each, 300 Hz:
These are raw USB stream figures. CSV files (text format) are larger per data point.
Troubleshooting¶
Radio and Connectivity¶
| Symptom | Likely cause | Resolution |
|---|---|---|
| No data after Start Receiver | Channel, radio mode or node count mismatch between nodes and receiver | Verify the settings on each node (nemi Sensor Config) and in nemi Studio / Web App — they must match exactly; wait up to one minute |
| Single node missing or appearing intermittently | Node number conflict or weak radio link | Re-program the nodes with unique numbers 1–8 (duplicates cause collisions); check the signal strength |
| Node shows short green blink every ~10 s | Node in sleep mode | Set the node to continuous or intermittent mode and send the settings |
| Node status yellow or red | Reception too weak or sampling rate too high for the radio mode | Shorten the distance, use a slower radio mode (500 / 125 kbit/s) or reduce the sampling rate |
| Low signal strength | Distance, obstruction or metal shielding | Reposition receiver antenna; switch to a slower radio mode |
| No reception after installing in a metal housing | Faraday cage effect | Route the antenna through a small opening; place the receiver outside the enclosure; extend the USB cable instead of the RF cable |
Missing Samples / Data Gaps¶
| Likely cause | Resolution |
|---|---|
| Weak radio link | Check signal strength; switch to a slower radio mode |
| Sensor buffering and replay | Expected behaviour: when the radio link is restored, the sensor retransmits buffered data alongside live data, temporarily increasing radio load |
| Radio channel overload | Reduce sampling rate or active channels, switch to a faster radio mode, or distribute nodes across multiple receivers; verify with the nemi Link 2400 Radio Channel Calculator |
| Host PC or Q.Station overload | In nemi Studio, watch USB Buffer and Message Queue in the status bar — rising values mean the computer cannot keep up; close other programs and unused plot windows |
USB / Software Issues¶
| Symptom | Likely cause | Resolution |
|---|---|---|
| No COM port is offered | Port list outdated or port held by another program | Click Refresh COM-Ports; re-plug the nemi Connect; close other programs that may hold the port |
| Receiver cannot be opened after several restarts | Stale USB device state | Unplug the nemi Connect, wait a few seconds and plug it in again; try a different USB port |
| COM port disappears / intermittent disconnects | USB hub with integrated Ethernet adapter | Connect directly to a PC USB port; if a hub is required, use a powered hub without Ethernet |
| nemi Studio Web App cannot connect | Browser without Web Serial (Safari, older Firefox) | Use Chrome or Edge; if reopening after the automatic reset fails, clear nemi Connect reset on connect and connect again |
| Sensor node cannot be configured via nemi Sensor Config | COM port still held after unplugging | Disconnect nemi Connect and the node; close and reopen nemi Sensor Config; then connect only the node |
| Embedded host does not boot correctly with the receiver attached | USB hub or Ethernet composite device conflict | Test without the hub or use a simpler USB topology |
Data and Timestamp Issues¶
| Symptom | Likely cause | Resolution |
|---|---|---|
| Timestamp jumps of exactly ≈ 71.6 minutes | Timer counter overflow in replayed (buffered) data | Correct algorithmically in post-processing |
| PC timestamps drift or step | Windows clock drift or NTP step correction | Keep the PC synchronized via NTP; for sub-millisecond multi-node timing in custom integrations, use the sensor timestamps together with the timing messages of the receiver (handled by the i4M Python API) |
| Data of several receivers does not align | Each receiver network has its own Stage 1 timing | Use one common host or NTP-synchronized PCs; for custom integrations, record the timing messages of each receiver and align in post-processing |
| Time base restarts | Receiver was restarted | When the receiver is (re)started its time base restarts; treat segments before and after a restart separately or align them by a known reference event |
Gantner Plugin Issues¶
| Symptom | Likely cause | Resolution |
|---|---|---|
| Missing or disappearing sensors | Plugin or controller overload | Reduce wired and wireless sampling rates; check Q.Station CPU load |
| Missing or disappearing sensors | Radio link unstable | Validate with nemi Connect + nemi Studio first; fix radio issues before returning to the Q.Station |
| No data | Incorrect channel / radio mode in plugin | Plugin configuration is independent of nemi Studio — verify channel and radio mode explicitly in the plugin |
| USB connection lost | — | Stop plugin → unplug nemi Connect → replug → restart plugin |
| Plugin crashes after approx. 71 minutes | 32-bit timer counter overflow not handled by older plugin versions | Update the nemione plugin to the current version (contact support) |
If nemi Studio is stable but the plugin is not, focus on controller load and plugin configuration. If nemi Studio also shows red node status, fix the radio network first.
Pre-Deployment Checklist¶
Complete this bench test before deploying sensors to difficult-to-access locations:
- All nodes powered and configured (correct channel, radio mode, node count, node number)
- nemi Studio shows all nodes with live data and green node status
- Signal strength acceptable for the planned deployment distance and environment
- Correct physical units and scaling verified for each channel
- Sleep / wake cycle tested (if used)
- CSV recording started; files created and spot-checked
- Working settings file saved; node map, firmware/software versions and receiver serial number documented
Product Changes¶
nemi Connect Firmware Updates¶
2023-12: nemi Link 2400 XR mode support New extended-range radio modes and a revised configuration message format were introduced. nemi Connect receivers with firmware dated before 2023-12-01 apply only the RF channel from the new configuration message and ignore the radio mode and network mode fields.
2025: nemi G+ ultra support nemi G+ ultra uses a new data message type that is not supported by earlier nemi Connect firmware. Use a current firmware version.
Firmware update procedure:
- Obtain the firmware binary from i4M support; provide your hardware serial number.
- Connect nemi Connect to the PC.
- Run the firmware update utility included with the firmware package.
- Follow the on-screen instructions. Do not unplug nemi Connect during flashing.
- After the update, verify the LED behaviour on reconnect (red flashing until the software connects).
nemi G+ ultra — Operation with nemi Connect¶
nemi G+ ultra transmits its high-rate data in a dedicated data message type; current nemi Connect firmware is required at all sampling rates.
| Sampling rate | Bandwidth (−3 dB) | Network constraint |
|---|---|---|
| 26.666 kHz | 6.72 kHz | One active axis; one node per network |
| 13.333 kHz | 4.66 kHz | Two axes with one node per network, or one active axis with up to three nodes |
| 6.666 kHz | 2.35 kHz | All axes with one node, or two axes with up to three nodes |
| 3.333 kHz | 1.17 kHz | All axes with up to three nodes, or one axis with up to eight nodes |
Values for 2 Mbit/s with a radio channel load below 70 %; check other combinations with the nemi Link 2400 Radio Channel Calculator.
See nemi G+ ultra for details.
Software Tools Version Reference¶
| Tool | Version reference |
|---|---|
| nemi Sensor Config | 2.1 (as of March 2024) |
| nemi Studio | 1.1 Beta 2 (as of November 2026) |
| nemi Studio Web App | 2026.11.04.3 (as of November 2026) |
| wirelessSensorPlot | 7.4 — legacy, no longer maintained; replaced by nemi Studio / nemi Studio Web App |
| nemi Link 2400 Radio Channel Calculator | Online tool, always current (replaces the Excel configuration tool) |
Checking Current Software Versions¶
- nemi Studio: title bar, top of the main window and About dialog (status bar, bottom right)
- nemi Studio Web App: version number next to the title; after an update, reload once without cache (Ctrl+F5)
- nemi Sensor Config: application title bar or About dialog
- nemione GI.bench plugin: plugin configuration dialog in GI.bench
- nemi Connect firmware: contact i4M support with the serial number of the receiver
Download links for the software are provided to customers; the software is not publicly available. Contact i4M support (info@nemi.one, +49 (0) 241 56 00 22 00) if your existing link does not resolve or provides an outdated version.
FAQ¶
Applications¶
When should I use nemi Connect vs. nemi Log vs. nemi EdgeBase? Use nemi Connect when a laptop, measurement PC or Gantner Q.Station is present at or near the measurement site. Use nemi Log for standalone SD-card measurements without a PC (optionally with cellular upload). Use nemi EdgeBase for autonomous long-term deployments with LTE/VPN connectivity, FTP download and MQTT streaming.
Can nemi Connect be used in wind turbine monitoring? Yes. Installations on hub, nacelle and blade components are practical. Closed metallic structures strongly attenuate 2.4 GHz radio — place the receiver inside the structure or route the antenna through a small opening. For setups combining high-rate nodes and longer-range nodes on the same turbine, use separate radio networks on different channels.
Can nemi Connect be used in ATEX / hazardous areas? Standard nemi Connect is not ATEX-certified. For potentially explosive atmospheres, consult i4M about certified barrier solutions. If explosion-proof (Ex d) enclosures are used, an antenna feedthrough is required to avoid the Faraday cage effect.
Can I use nemi Connect for inclination / tilt monitoring? Yes, with suitable sensor nodes. nemi G+ derives static pitch and roll from the accelerometer gravity vector; nemi Studio shows roll, pitch and yaw live in the Attitude tab. For inclination-only applications, use the smallest range (±2 g) and a low sampling rate (e.g. 125 Hz) to maximize resolution and minimize noise.
Can acceleration be integrated to obtain velocity or displacement? Double integration of acceleration is impractical for absolute displacement measurement — offsets on the accelerometer cause large position drift. For displacement or velocity, use dedicated sensor principles (encoders, LVDTs, potentiometric displacement sensors) connected to nemi DAQ analog inputs.
Can nemi Connect be used in underwater or water-adjacent applications? nemi Connect itself (IP41) is not rated for water immersion. House it in a sealed enclosure and extend the USB cable. 2.4 GHz radio does not propagate through water — sensor node antennas must be above the waterline.
Connectable Sensors¶
Which sensor nodes are compatible with nemi Connect? All nemi Link 2400 nodes: nemi G+, nemi G+ nano, nemi G+ ultra, nemi G+ ultra nano, nemi DAQ, nemi DAQ nano, nemi Struct and nemi Motion.
Can I connect external sensors (strain gauges, Pt100, 0–10 V, geophones, IEPE accelerometers)? Not to nemi Connect directly — external transducers are connected to a nemi DAQ or nemi DAQ nano node, which transmits the data wirelessly to nemi Connect. The DAQ input configuration (e.g. strain gauge, Pt100/Pt1000 or 0–10 V) must be specified at order time; reference voltage sensors (e.g. strain gauges) and reference current sensors (e.g. Pt100/Pt1000) cannot be combined on one nemi DAQ. Geophones and IEPE accelerometers require external signal conditioning.
Can the gyroscope of a nemi G+ measure shaft speed? Yes. The gyroscope measures rotational velocity with ranges up to ±4,000 °/s (≈ 666 rpm). nemi Studio provides a Rotation Angle tab with rotation angle, speed (rpm) and revolution count for sensors mounted on rotating shafts.
Power Supply¶
What power supply does nemi Connect require? 5 V via USB-C. The device is powered entirely by the USB connection — no separate power supply is needed.
Does nemi Connect have a battery? No. nemi Connect is powered by USB; battery life considerations apply only to the wireless sensor nodes.
Can I power nemi Connect from a USB hub? Yes, if the hub is stable and provides enough power. Direct connection to the PC is preferred during commissioning; avoid untested hubs, especially hubs with integrated Ethernet adapters, in critical installations.
How does the configuration affect sensor battery life? Higher sampling rates and slower radio modes (longer on-air time) increase power consumption. Using intermittent operating mode, sleep mode and disabling unused IMU channels extend battery runtime. See the individual sensor node pages for runtime estimates.
Radio and Range¶
What is the operating range of the nemi Link 2400 radio? Up to 350 m, strongly environment-dependent. Indicative figures (indoor / urban): ~20 / 50 m at 2 Mbit/s, ~50 / 125 m at 1 Mbit/s, ~90 / 225 m at 500 kbit/s, ~140 / 350 m at 125 kbit/s. Metal structures, concrete and wet materials reduce range significantly.
How can I extend radio range? Switch to a slower radio mode (500 kbit/s or 125 kbit/s), ensure line of sight between receiver and sensors, elevate the receiver antenna and avoid placing the receiver inside metal enclosures. Extend the USB cable rather than the antenna cable.
How many sensors can one nemi Connect receive at once? Up to 8 wireless sensor nodes. All must share the same radio channel, radio mode and node count.
Can one nemi Connect receive more than eight nodes? No. Use multiple receivers on different radio channels when more than eight nodes are required or when the channel load is too high (check with the nemi Link 2400 Radio Channel Calculator). Each receiver appears as a separate COM port; several receivers can be connected to one PC.
Which radio channel should I start with? Channel 0 is the default and a good choice. For parallel networks use channels 0, 25, 50 and 75. Channel 10 also works but may be slightly more exposed to Wi-Fi interference than channel 0.
Can I run two independent nemi Connect networks at the same location? Yes. Assign each a different channel (e.g. channel 0 and channel 25). Each receiver connects to its own COM port independently.
Integration¶
Does nemi Connect have a native Ethernet or Profinet port? No. The only physical interface is USB-C. For Ethernet or fieldbus output, use a Gantner Q.Station as intermediary, or build a Python-based bridge using the Python API.
Can I use nemi Connect from Linux or macOS? Yes. The nemi Studio Web App runs in Chrome or Edge on any operating system that supports Web Serial, and the Python API is platform-independent. nemi Studio (desktop) is available for Windows 10 or higher; Linux or macOS versions can be implemented on request.
How precise is multi-sensor timing for structural dynamics analysis? Within one nemi Link 2400 network, the sensor clocks are synchronized to the receiver with ±10 µs with a stable radio connection — suitable for modal analysis, structural dynamics and waveform correlation across channels.
How are several receivers aligned in time? Sensor nodes are synchronized to their own receiver first (Stage 1). Receiver-to-receiver alignment is handled downstream: via one common host PC, NTP-synchronized PCs, Q.Station timing, the timing messages of the receivers or post-processing.
Can a nemi Log be used instead of a nemi Connect? Yes. A nemi Log can be used as a USB receiver in the same way as a nemi Connect, e.g. with nemi Studio or the nemi Studio Web App.
Why does my data have timestamps that jump by exactly ~71.6 minutes? This is the 32-bit timer counter overflow in replayed (buffered) data. See Timer Counter Overflow in the Integration section.
Can I integrate nemi Connect into a Gantner Instruments system? Yes. Connect nemi Connect via USB to a Q.Station and use the GI nemione plugin; the data then appears in GI.bench and GI.cloud. Several nemi Connect units per Q.Station are possible.