nemi Studio¶

Description¶
nemi Studio is the Windows desktop software for the configuration, live display, analysis and recording of nemione® wireless sensor nodes and telemetry modules (e.g. nemi G+, nemi DAQ) as well as OEM-specific i4M wireless sensor nodes in the nemi Link 2400 wireless network.
nemi Studio is the successor of the legacy software wirelessSensorPlot. It works either with a nemi Connect USB receiver (or a nemi Log in USB receiver mode) or with a nemi EdgeBase via a network connection. Typical tasks are:
- Commissioning a sensor network: setting up the receiver, checking signal strength, latency and sampling rates of every node.
- Configuring the runtime measurement settings of each node wirelessly (mode, sampling rate, measuring ranges, IMU ranges).
- Live display of up to eight nodes per radio channel, with large single-node plots and a common plot for comparing synchronised sensors.
- Live analysis: FFT, attitude (roll, pitch, yaw with automatic compass calibration) and rotation angle of sensors on rotating shafts.
- Time-synchronised CSV recording, optionally resampled into one common file for all nodes.
For a browser-based alternative without installation, see nemi Studio Web App.
Typical Applications¶
- Commissioning and range checks — verify radio parameters, signal strength and data completeness before a sensor is mounted at its final location.
- Short- and medium-duration test measurements on a measuring computer — vibration tests, ride and vehicle measurements, structural tests.
- Rotating machinery — rotation angle, speed and revolutions of sensors mounted on shafts, including compensation of centrifugal acceleration.
- Orientation measurements — roll, pitch and yaw via sensor fusion of accelerometer, gyroscope and magnetometer.
- Acoustic inspection — making one accelerometer axis audible (e.g. bearing or gear noise) and saving it as a WAV file.
- Live monitoring of a nemi EdgeBase installation over VPN or local Ethernet (EdgeBase
logging_mode 6).
Technical Specifications¶
System Requirements¶
| Parameter | Value |
|---|---|
| Operating system | Windows 10 or higher (Linux or macOS versions on request) |
| Installation | None — unzip the ZIP package provided by i4M and run nemiStudio.exe |
| Folder | Folder with write access (e.g. below Documents, not C:\Program Files) |
| Receiver connection | USB (nemi Connect, virtual COM port) or TCP/IP (nemi EdgeBase, port 9413) |
| Current version | 1.1 Beta 2 (November 2026) |
Supported Devices¶
| Device | Description |
|---|---|
| nemi Connect | USB receiver for the nemi Link 2400 network, connected to the measuring computer via USB (virtual COM port) |
| nemi Log (USB receiver mode) | Can be used like a nemi Connect as a USB receiver |
| nemi EdgeBase | Edge computer with integrated receiver; provides the live data via TCP/IP (logging_mode 6, port 9413) |
| nemi G+ / nemi G+ nano | Wireless accelerometer (M08: ±2 / 4 / 8 g, M40: ±10 / 20 / 40 g), sampling rate up to 4,000 Hz, integrated IMU |
| nemi G+ M256 | Like nemi G+, with extended ranges ±16 / 32 / 64 / 128 / 256 g of the main accelerometer |
| nemi G+ ultra / nemi G+ ultra nano | Wireless vibration sensor (up to 26.7 kHz, ±2 … 16 g, axes individually switchable), integrated IMU |
| nemi DAQ / nemi DAQ nano | Telemetry module with up to three analog channels (strain gauges, voltages, Pt100/Pt1000, 0–10 V), up to six digital inputs and an integrated IMU |
Radio Data Rates¶
| Radio data rate | Typical range (indoor / urban) | Max. nodes per channel |
|---|---|---|
| HS 2 Mbit/s (High Speed) | ~20 m / ~50 m | 1 (Low Latency), 3 or 8 |
| HS 1 Mbit/s (High Speed) | ~50 m / ~125 m | 1 (Low Latency), 3 or 8 |
| XR 500 kbit/s (Extended Range) | ~90 m / ~225 m | 1 (Low Latency), 3 or 8 |
| XR 125 kbit/s (Extended Range) | ~140 m / ~350 m | 1 (Low Latency), 3 or 8 |
The maximum possible sampling rates depend on the radio data rate and the number of nodes per channel; nemi Studio pre-selects suitable default sampling rates when these settings are changed. To plan a configuration in advance, check the radio channel load with the nemi Link 2400 Radio Channel Calculator. Range always has to be validated on site — see nemi Link 2400.
nemi Studio vs. nemi Studio Web App¶
| Feature | nemi Studio | nemi Studio Web App |
|---|---|---|
| Platform | Windows 10+ desktop application | Browser (Chrome / Edge recommended; Firefox, Safari with restrictions) |
| Installation | Unzip ZIP package, run nemiStudio.exe |
None — open https://nemi.one/studio/ or the address of the EdgeBase |
| nemi Connect (USB) | Yes (virtual COM port) | Yes, in browsers with Web Serial (Chrome, Edge, newer Firefox) |
| nemi EdgeBase (network) | Yes (logging_mode 6, port 9413) |
Yes — the app is served by the EdgeBase itself; works in every current browser incl. iPad. EdgeBase units produced before November 2026 need a software update, which i4M installs remotely on request. |
| Live display | 3 node columns (any of 8 nodes selectable), Big Plot, Common Plot | Up to 8 sensor columns side by side, magnified view |
| Analysis | FFT, Attitude (3D), Rotation Angle, Sound Output | FFT, Waterfall 2D/3D, Tilt Measurement, Orientation 2D/3D |
| nemi G+ M256 extended ranges | Yes (16 … 256 g) | Not listed in the current manual |
| Zero / scaling | Node Zero and Scaling (offset and scale for every channel) | Zero for nemi DAQ channels and tilt angle |
| Recording | One CSV per node and data stream; optional resampled common file | One common CSV file on a common time grid |
| Settings storage | INI files, last_session_settings.ini |
Locally in the browser |
Migration from wirelessSensorPlot¶
wirelessSensorPlot is no longer maintained. Former users will find the familiar workflow in nemi Studio:
- The structure with Step 1 (connection) and Step 2 (receiver settings), three node columns with node selection, Update Node Settings and Save / Load Settings to/from File has been kept.
- The CSV output is split into IMU, vibration, telemetry, digital-input, timing and debug files with the same time columns (
Timestamp (ASCII),Timestamp (s),Timecounter (s)). - New: live FFT, attitude and rotation-angle analysis, Big Plot, Common Plot, Node Zero and Scaling, Sound Output, resampling into a common file, support of nemi G+ M256.
- Persistent radio parameters of the sensor nodes are still written with nemi Sensor Config via USB.
First Commissioning¶
Installation¶
- Unzip the ZIP package provided by i4M technologies into a folder with write access (e.g. below Documents).
- Run
nemiStudio.exe. If error messages appear, contact your i4M contact person. - When settings of a previous session are found, nemi Studio asks whether they should be loaded. Choose Yes to continue with the previous configuration (node types, settings, node names, compass calibrations, rotation axes) or No to start with the default settings.
Depending on the system settings, administrator rights may be required to record data into certain folders.
Prerequisites¶
- The sensor nodes have been configured once with nemi Sensor Config via USB (radio channel, radio data rate, node number, number of nodes per channel). See the respective sensor page, e.g. nemi G+.
- A nemi Connect with a data-capable USB cable, or a nemi EdgeBase reachable via VPN or local network.
Connecting a nemi Connect (USB)¶

- Disconnect the nemi Connect from the USB port of the measuring computer.
- Start nemi Studio.
- Plug the nemi Connect into a USB port. Windows plays the sound for a new USB device and the LED of the nemi Connect flashes red slowly.
- nemi Studio detects the new nemi Connect and automatically selects the corresponding COM port. A message is shown in the Debug Console.
- Click Connect COM-Port. Messages of the receiver appear in the Debug Console.
- Wait until the message "Waiting for Config Messages to start …" appears.
- Optional: load previously saved settings with Load Settings from File.
- Select the radio data rate and the number of sensor nodes per radio channel (Step 2). These settings must match the settings of the sensor nodes (set with nemi Sensor Config). Except for the radio channel, they can only be changed after Disconnect COM-Port.
- Set the radio channel.
- Select the module type and the settings of each sensor node, unless they were loaded from a file.
- Click Start Receiver. The receiver is configured and the settings are sent to all sensor nodes of the network.
- Switch on the sensor nodes on the configured channel if not already done.
- Measurement data is now displayed in the live plots and the nemi Connect flashes green.
Note
The receiver only starts when configuration messages for all nodes of the network have been sent — Start Receiver does this automatically. When the receiver is (re)started, its time base restarts; all displays continue automatically.
Connecting a nemi EdgeBase (Network)¶
A nemi EdgeBase can provide its live data via a VPN connection (mobile network) or a local Ethernet connection. For this, logging_mode 6 must be set in the sensor_config.yml of the EdgeBase; the EdgeBase then provides all measurement data on TCP/IP port 9413. All functions of nemi Studio including recording are available.

- Start nemi Studio and click Connect EdgeBase.
- Enter the IP address of the nemi EdgeBase (default
192.168.101.2) and leave the port at9413. The computer must be able to reach the EdgeBase via VPN or the local network. - Click Connect to nemi EdgeBase.
- Wait until "Waiting for Config Messages to start …" appears in the Debug Console.
- Set up the radio network and the nodes as described for the nemi Connect and click Start Receiver.
- To disconnect, click Disconnect EdgeBase.
The internal IMU of the EdgeBase can be displayed in any node column by selecting EdgeBase IMU in the node selection; its data is shown in the IMU plot. The EdgeBase IMU uses fixed settings: 416 Hz, ±4 g, ±500 °/s, ±8 gauss.
First Measurement — Checklist¶
- Radio channel, data rate and number of nodes in nemi Studio match the settings of the sensor nodes.
- Each node has a unique node number on the channel.
- All nodes show green in the Node Status table, plausible signal strength and low latency.
- Wait approx. 10 s after starting the receiver until the time synchronisation is complete and Start Recording is enabled.
- Make a short test recording and check the files before the actual measurement.
User Interface¶
The main window consists of a header area with the global functions and three node columns below, each of which displays and configures one sensor node.
| Area | Function |
|---|---|
| Step 1 | Connection: select the COM port of the nemi Connect and Connect COM-Port, refresh the port list, or Connect EdgeBase. The frame turns green as soon as the connection is established. |
| Step 2 | Receiver: radio data rate, number of sensor nodes per channel, radio channel, Load Settings from File, Start Receiver and Change Channel. |
| Recording | Start Recording, recording options and Recorded Channels. |
| Node Settings | Update Node Settings, Save Settings to File, Edit Node Names, Node Zero and Scaling. |
| Plot Control | Common Plot, Pause Plotting, Clear Plots and Consoles. |
| Node Status | Overview of up to 8 nodes: signal strength and total sampling rate per node, colour-coded. |
| Debug Console | Status and error messages of the system and the receiver; below: Theme (Bright / Dark), Copy Console and Clear Console. |
| Node columns | Node selection, module type, Sensor Configuration, buttons Node Details, Big Plot, Live Analysis, Sound Output, checkbox Log Angular Data, status line, live plot of the primary sensor (top), IMU plot (bottom) and node console. |
| Status bar | Bottom right: USB Buffer and Message Queue, Sync Info and About. |
If eight sensor nodes per radio channel are used, any of the eight nodes can be shown in each of the three columns via the node selection at the top left of the column. The settings of a node are kept when another node is selected and restored when it is selected again.
Detailed Configuration¶
Configuration Layers¶
| Layer | Tool | Settings |
|---|---|---|
| Persistent radio parameters (stored in the node) | nemi Sensor Config v2.1+ via USB | Radio channel, radio data rate, node number, max. nodes per channel |
| Receiver settings | nemi Studio, Step 2 | Radio data rate, nodes per channel, radio channel — must match the nodes |
| Runtime measurement settings (sent wirelessly) | nemi Studio, node columns | Mode, sampling rate, measuring ranges, IMU ranges, gains, digital inputs |
Changing Radio Settings During Operation¶
- Radio channel: after Start Receiver, the channel can be changed at any time with Change Channel; the change takes effect immediately.
- Data rate and number of nodes: click Disconnect COM-Port, wait until the nemi Connect flashes red again, click Connect COM-Port, change the settings and click Start Receiver.
Module Type and Automatic Detection¶
Select the type of the sensor node in the module type list at the top right of a column: nemi G+, nemi G+ nano (also for the nemi G+ M256), nemi DAQ, nemi DAQ nano or nemi G+ ultra, nemi G+ ultra nano. The settings area adapts to the selected type.
If the selected type does not match the node transmitting on this number, nemi Studio switches the column to the correct type automatically as soon as the first data message arrives and sends the stored settings to the node. During the first seconds the node may still use other settings (e.g. a nemi G+ transmits 4,000 Hz for approx. 10 s).
Node Settings¶
| Setting | Sensor type | Options / meaning |
|---|---|---|
| Mode | all | 0 = Sleep, 1 = Continuous, 2 = Intermittent (measure for Sample [s], then sleep for Sleep [s]) |
| ODR | nemi G+ | 4,000 / 2,000 / 1,000 / 500 / 250 / 125 Hz |
| Range | nemi G+ | 2 g (M08/M256) / 10 g (M40), 4 g (M08/M256) / 20 g (M40), 8 g (M08/M256) / 40 g (M40); 16, 32, 64, 128, 256 g (M256 only) |
| ODR | nemi G+ ultra | 26,667 / 13,333 / 6,666 / 3,333 Hz |
| Range / X-, Y-, Z-Axis | nemi G+ ultra | 2, 4, 8, 16 g; each axis can be switched on or off |
| AF ODR | nemi DAQ | Sampling rate of the analog channels; the available values depend on the number of active channels; LP = low-power mode |
| Ch. 1 / Ch. 2 / Ch. 3/4 Gain | nemi DAQ | off, 1×, 4×, 8×, 16×, 32×, 64×, 128×; channels with gain off are switched off. Channel 4 only exists on customer-specific four-channel modules and always uses the gain of channel 3. |
| IMU ODR | nemi DAQ | 52 … 3,332 Hz (standard IMU) / 60 … 3,840 Hz (High-G IMU) |
| GPIO | nemi DAQ | High level of the digital inputs: off, 5 V, 12 V, 24 V. The 5 V setting is also used for 3.3 V signals (same threshold). |
| IMU ACC Range | all | off, 2, 4, 8, 16 g; 32 … 256 g only for nodes with High-G IMU |
| IMU GYR Range | all | off, 125, 245, 500, 1,000, 2,000, 4,000 °/s |
| IMU MAG Range | all | off, 4, 8, 12, 16 gauss |
The nemi G+ M256 transmits the same data messages as a nemi G+ and is configured as nemi G+. Its integrated IMU does not support the extended ranges. In the ranges from 32 g the main sensor uses a separate high-g element whose zero offset can be several tenths of a g; use Node Zero and Scaling if required.
Extended ranges. The ranges marked (M256) (16 … 256 g) and the IMU ACC ranges 32 … 256 g are greyed out by default. Each sensor node reports its hardware in the node console when it starts (when it is switched on and when it applies new settings). nemi Studio evaluates these messages and enables the extended ranges for this node:
| Hardware reported by the node | Enabled ranges |
|---|---|
| High-g main accelerometer (nemi G+ M256) | nemi G+ ranges 16 … 256 g (M256) |
| High-g IMU | IMU ACC Range 32 … 256 g |
The detection is stored per node number in last_session_settings.ini. If a node reports a different accelerometer or IMU, the extended ranges are disabled again. If a loaded settings file contains an extended range for a node without this detection, the range is reduced to the largest standard range (main sensor 8 g, IMU 16 g); the Debug Console shows a message and Update Node Settings turns red.
Sending Settings (Update Node Settings)¶
When a setting is changed, the text of Update Node Settings turns red: the changes have not yet been sent. A click sends the settings of all nodes. A node only changes its configuration if the new settings differ from the current ones; it may take up to approx. one minute until a node applies new settings.
Node Names¶

Click Edit Node Names to name the up to eight nodes and confirm with Set Sensor-Node Names. The names are shown in the node selection and are used in the file names of recordings.
Saving and Loading Settings¶
Save Settings to File stores all settings in an INI file (.ini) that can also be edited with a text editor; Load Settings from File loads them again. The file contains the radio settings, all node settings, node names, zero/scaling values, recording channels, the compass calibrations and rotation axes. When the program is closed, the current settings are additionally saved as last_session_settings.ini next to the program.
If a loaded file specifies a different module type than the node actually connected, the type is corrected automatically when the first data of this node arrives.
Node Zero and Scaling¶

Node Zero and Scaling opens a table per node with the raw value, offset, scale and scaled value of each channel (primary sensor, analog channels, IMU). The displayed and recorded value is:
Zero averages the signal for 3 seconds and sets the offset so that the current value becomes zero. Apply takes over the values; they are stored with the settings.
Live Data & Analysis¶
Live Plots and Channel Selection¶
The upper plot of a column shows the primary sensor (accelerometer or analog channels), the lower plot the IMU of the node. The last 10 seconds are displayed. The checkboxes above the plots select the displayed channels; the list next to the IMU plot selects accelerometer, gyroscope, magnetometer or digital inputs (Plot DIN, nemi DAQ). Pause Plotting freezes the plots, Clear Plots and Consoles resets plots, consoles and status displays.
The display unit of accelerometers is always g (1 g = 9.81 m/s²). For the nemi DAQ, the unit of all analog channels is selected with Y Axis Unit:
| Unit | Conversion |
|---|---|
| mV/V | Default unit, ratiometric to the ADC reference / supply voltage of the nemi DAQ (2.5 V) |
| V | mV/V multiplied by the reference voltage of 2.5 V |
| °C (Pt 100) / °C (Pt 1000) | Direct conversion for Pt100 / Pt1000 resistance thermometers supplied with constant current |
| 0 - 10 V | For 0–10 V sensors; the signal attenuation of the 0–10 V mode is taken into account (the nemi DAQ must be configured accordingly) |
Status Displays¶
| Display | Meaning |
|---|---|
| Temperature | Temperature of the node |
| Voltage | Battery voltage: 4.1 V ≈ 100 % charge, approx. 3.3 V ≈ below 20 % |
| Signal | Signal strength, converted from the RSSI value into a distance relative to the maximum range of the selected radio mode (linear scale) |
| Latency (ms) | Age of the latest data packet; indicates whether the node is live |
| Hz fields | Live sampling rates of the primary sensor and of the IMU / digital inputs, monitored continuously |
| Node Status table | Green: all data streams complete. Yellow: the last sample is older than 2 s (the node may deliver buffered data later, or is in sleep mode). Red: at least one sample was lost within the last second — check the reception or reduce the sampling rate. |
| USB Buffer (status bar) | Bytes received from the nemi Connect (or the EdgeBase) that nemi Studio has not yet read. Normally close to 0; a rising value means that the computer cannot keep up with the data stream. |
| Message Queue (status bar) | Decoded sensor messages waiting for display and recording at the last screen update (every 50 ms). A permanently high or rising value means that display and recording cannot keep up. |
Big Plot¶

Big Plot opens a large plot of one node with all channels: primary sensor, IMU (acceleration, angular rate, magnetic field) and system data (temperature, voltage, RSSI). Options: time window, pause, legend, dual y-axis (IMU on the right axis), OpenGL rendering, automatic or manual scaling, Clear and Save PNG….
Common Plot¶

Common Plot shows channels of all eight nodes in one plot, e.g. to compare synchronised sensors. Select the channels and their y-axis (left/right) in the tabs Node 1 … Node 8; the controls are the same as in the Big Plot. The selection is saved with the settings.
Node Details¶

Node Details shows information of a node (type, firmware, radio mode, temperature, voltage) and the quality of the time synchronisation over the last 100 synchronisation events (synchronisation accuracy and round-trip time).
Sound Output¶

Sound Output (nemi G+ and nemi G+ ultra) makes one axis of the accelerometer audible on the loudspeaker of the computer, e.g. to listen to bearing or gear noise. A high-pass filter removes the static part, an optional low-pass limits the bandwidth; Max Signal until Clipping and Gain set the volume. Record to WAV saves the signal as an audio file with the sampling rate of the sensor.
Consoles¶
The node console at the bottom of each column shows the messages of the node (start-up messages, synchronisation, settings); the Debug Console at the top right shows the messages of the receiver and of nemi Studio. Copy to Clipboard (node console) and Copy Console (Debug Console) copy the complete content to the clipboard, e.g. to paste it into an e-mail to i4M.
Live Analysis: FFT¶
Live Analysis opens the analysis window of a node with the tabs FFT, Attitude and Rotation Angle. Attitude and rotation angle are calculated continuously for every node with IMU data — also when the window is closed — so that they can be recorded at any time.

The FFT tab shows the amplitude spectrum of the selected channels (primary sensor, analog channels, IMU). Settings: frequency and magnitude range, window length in seconds, overlap of successive FFTs, window function (Hanning, Hamming, Blackman, Flat Top) and linear or logarithmic (dB) display. The amplitude is corrected for the window function, i.e. a sine with an amplitude of 1 g is shown as 1 g (0 dB). The sampling rate of each channel is measured automatically.
Live Analysis: Attitude¶

The Attitude tab shows the orientation of the node in space as roll, pitch and yaw angles (Euler angles ZYX, yaw 0 … 360°) and quaternion, as time plot and as 3D view with the product model. The attitude is calculated by sensor fusion of gyroscope, accelerometer and magnetometer:
- Roll and pitch are derived from gravity and are available within a second after the start.
- Yaw (heading) requires a calibrated compass. The magnetometer is calibrated automatically while the sensor is moved: rotating it once flat on a table gives a planar calibration, tilting it in different directions a full 3D calibration. The status line shows the progress. The calibration is stored per node.
- Magnetic disturbances (e.g. steel, magnets, mobile phones) are detected; the compass is then ignored until the field is plausible again. Without compass (also with IMU MAG Range = off) the yaw angle drifts slowly.
- The gyroscope offset is learned automatically, in particular while the sensor is at rest.
Controls: Use magnetometer switches the compass on or off; Zero heading defines the current horizontal direction of the sensor X axis as 0° (north); Clear zero returns to magnetic north. 3D view: drag with the mouse to rotate, mouse wheel to zoom, double-click to reset.
Note
The attitude estimation assumes that gravity dominates the measured acceleration. It is not suitable for sensors on fast-rotating shafts (centrifugal acceleration) — use the Rotation Angle tab for this purpose. North means magnetic north; there is no declination correction.
Live Analysis: Rotation Angle¶

The Rotation Angle tab provides the rotation angle of a sensor mounted on a rotating shaft, together with the speed (rpm) and the number of revolutions. The angle is determined from the gyroscope and continuously corrected by the direction of gravity. Constant accelerations in the sensor (centrifugal acceleration of an eccentric mounting position, accelerometer offsets) and tangential accelerations during run-up are estimated and removed automatically; the radius of the mounting position does not need to be known.
- Rotation axis: detected automatically after approx. two revolutions and stored per node. Until then, the standard mounting is assumed (housing bottom on the shaft surface, X along the shaft, Z radially outwards). Reset axis forgets the axis, e.g. after mounting the sensor on another shaft.
- Zero point: by default 0° is the position in which the housing top (+Z) points upwards. Zero here defines the current position as 0°, Reset zero (sensor up) returns to the default.
- Revolutions: counted continuously with sign; Reset revolutions sets the counter to zero.
- Status: detected axis, axis misalignment, shaft inclination, zero reference, used accelerometer axes, gyroscope saturation, lever arm, phase residual and angle uncertainty.
Recommendations and limits:
- Use a high IMU sampling rate (nemi DAQ: e.g. 3,332 Hz) and a gyroscope range matching the speed (4,000 °/s ≈ 666 rpm). At least approx. 8 samples per revolution are required for good accuracy.
- Above the gyroscope range, the angle is tracked from the accelerometer only, with reduced accuracy. If the radial accelerometer axis saturates due to centrifugal acceleration, the tangential axis is used alone.
- The shaft must not be vertical.
Recording Data¶
Starting and Stopping a Recording¶
After connecting and starting the receiver, nemi Studio synchronises the sensor clocks with the clock of the computer. Recording is possible as soon as Start Recording is enabled (approx. 10 s).
- Select the recording options below Start Recording (see table).
- Optional: select the channels to be recorded with Recorded Channels.
- Click Start Recording and select the folder for the files (the last folder is suggested).
- Click Stop Recording to finish the recording. Files without data are deleted automatically.
| Option | Meaning |
|---|---|
| Log full Timestamps | Write UNIX and ASCII timestamps in addition to the time counter |
| Log Debug only | Only record debug data (console output), no measurement data |
| Record Sync | Write an additional file with time synchronisation data |
| Resample Data / Output Rate | Create a unified CSV file with all nodes at a common sampling rate after the recording |
| Del. Orig. Data | Delete the original node files after successful resampling |
| Log Angular Data (per node) | Add attitude and rotation angle columns to the IMU file of the node. Requires IMU ACC and GYR; MAG is optional |

In the Recorded Channels dialog the channels of each node to be written are selected; the dialog also shows live values. The selection cannot be changed during a running recording.
Files and Format¶
The files are written as CSV with semicolon (;) as separator; the first line contains the column names. The file name starts with date and time of the start of the recording and contains the node name:
| File | Content |
|---|---|
<datetime>_<node>_imu_data.csv |
IMU of the node (ACC / GYR / MAG), temperature, battery voltage, RSSI; optionally angle columns |
<datetime>_<node>_vibration_data.csv |
nemi G+ / G+ ultra: primary accelerometer in g (incl. zero and scaling) |
<datetime>_<node>_telemetry_data.csv |
nemi DAQ: analog channels in the selected unit (incl. zero and scaling) |
<datetime>_<node>_din_data.csv |
nemi DAQ: digital inputs |
<datetime>_timing_data.csv |
Time synchronisation data (only with Record Sync) |
<datetime>_debug.csv |
Debug data (all console messages) — can be sent to i4M for troubleshooting |
<datetime>_unified_resampled_<N>Hz.csv |
Resampled common file of all nodes (only with Resample Data) |
Each sample has three time columns:
| Column | Content |
|---|---|
Timestamp (ASCII) |
Human-readable date and time (local time) with microseconds |
Timestamp (s) |
UNIX time in seconds since 1 January 1970, microsecond resolution |
Timecounter (s) |
Running time of the measurement, set to 0 at the start of the recording |
Angular Data¶
If Log Angular Data is enabled for a node, the IMU file additionally contains the columns Roll (deg), Pitch (deg), Yaw (deg), Quat W … Quat Z (attitude, as displayed, including Zero heading) as well as Rotation Angle (deg), Rotation Speed (rpm) and Revolutions. Values that are not yet valid (e.g. before the rotation axis is known) are left empty. With IMU MAG Range = off the fusion works with accelerometer and gyroscope only: roll, pitch and the rotation angle are unaffected, the yaw angle is then based on the gyroscope only and drifts slowly.
Resampling¶
With Resample Data all recorded files are combined into one CSV file with the selected output rate after the recording. The signals are low-pass filtered (anti-aliasing) and interpolated onto the common time grid. Gaps in the data remain empty in all channels; digital inputs keep their 0/1 values; angle columns are processed correctly across the 0°/360° transition. The progress is shown in a separate window and can be cancelled.
Viewing Recorded Data¶
For a quick look at the data, the free FlowCSV Viewer can open the files directly. For further analysis, e.g. GNU Octave, MATLAB or Python (pandas, NumPy) can be used.
Time Synchronisation¶
The sensor nodes are time-synchronised within the radio network (±10 µs with a stable radio connection). nemi Studio additionally relates the time of the receiver to the clock of the computer and assigns a synchronised timestamp to each sample. The computer clock should therefore be synchronised with a time server (NTP, Windows time settings). Several receivers on one computer, or several computers synchronised to the same NTP server, can thus record time-synchronously.

Sync Info (status bar, bottom right) shows the quality of the synchronisation: the change of the offset between PC and receiver time, the estimated clock rate deviation in ppm and the model error. It is mainly intended for diagnosis.
Troubleshooting¶
| Problem | Possible cause / solution |
|---|---|
| No COM port is offered | Click Refresh COM-Ports; re-plug the nemi Connect; check whether another program uses the port; use a data-capable USB cable. |
| The receiver cannot be opened after several restarts | Unplug the nemi Connect, wait a few seconds and plug it in again. |
| No data after Start Receiver | Check radio channel, data rate and number of nodes (must match the nodes); switch the nodes on; wait up to one minute. |
| Node Status yellow or red | Reception too weak or sampling rate too high for the radio mode: shorten the distance, use an XR mode or reduce the sampling rate. |
| Start Recording stays disabled | The time synchronisation is not yet complete — wait approx. 10 s after the start of the receiver. |
| Compass remains calibrating | Rotate the sensor once completely (flat), better also tilt it; avoid steel parts and magnets nearby. |
| Rotation angle: axis nearly vertical | The assumed or detected axis is vertical; rotate the shaft to detect the real axis. Vertical shafts are not supported. |
| Files cannot be written | Choose a folder with write access or start nemi Studio with administrator rights. |
| USB Buffer / Message Queue keep rising | The computer is overloaded: close other programs, close unused plot windows or use another computer. |
| Range 16 … 256 g (M256) or IMU ACC 32 … 256 g greyed out | The node has not (yet) reported the matching hardware: switch the node off and on again while nemi Studio is connected. |
| Node settings not applied | Check that Update Node Settings was clicked (text no longer red); a node may need up to approx. one minute to apply new settings. |
For further questions, please contact your i4M contact person or info@nemi.one, phone +49 (0) 241 56 00 22 00. When reporting a problem, a recording with Log Debug only or the _debug.csv file of the measurement helps with the analysis.
Product Changes¶
Version Information¶
| Software | Version | Status |
|---|---|---|
| nemi Studio | 1.1 Beta 2 | Current (November 2026) |
| nemi Studio | 1.0 | February 2026 |
| wirelessSensorPlot | 7.4 | Legacy, no longer maintained — replaced by nemi Studio / nemi Studio Web App |
The version is shown in the title bar, at the top of the main window and in the About dialog (status bar, bottom right).
What Is New in Version 1.1¶
- New sensor fusion for the attitude (roll, pitch, yaw) with automatic online calibration of the magnetometer — no calibration procedure required.
- Rotation Angle: clean rotation angle, speed and revolutions of sensors mounted on rotating shafts, including compensation of centrifugal acceleration.
- 3D view with product models of nemi G+, nemi G+ ultra and nemi DAQ and smooth (interpolated) motion.
- Support of the nemi G+ M256 with measuring ranges up to ±256 g. Extended ranges are only offered when the node reports the matching hardware (nemi G+ M256 or High-G IMU).
- Log Angular Data also works without magnetometer (IMU MAG Range = off).
- Consoles can be copied to the clipboard and cleared; status bar with USB Buffer, Message Queue, Sync Info and About.
- IMU gyroscope ranges are listed in ascending order; magnetometer axes of the nemi G+ ultra are aligned with nemi G+ and nemi DAQ.
- Recording: angle columns (attitude and rotation angle), resampling keeps data gaps empty, many stability improvements.
Licenses¶
nemi Studio is proprietary software of i4M technologies GmbH. It uses the following open-source components; their license texts are contained in the file THIRD_PARTY_LICENSES.txt and in the About dialog:
- Qt 5 (The Qt Company) — GNU Lesser General Public License v3
- Qwt 6.3 — Qwt License v1.0 (based on LGPL v2.1)
- cgltf — MIT License
FAQ¶
General¶
Does nemi Studio need to be installed?
No. Unzip the ZIP package provided by i4M into a folder with write access and run nemiStudio.exe.
Which operating systems are supported? Windows 10 or higher. Versions for Linux or macOS can be implemented on request. Alternatively, the nemi Studio Web App runs in a browser on any operating system.
How do I know which version I am running? The version is shown in the title bar, at the top of the main window and in the About dialog.
I used wirelessSensorPlot before — what changes? The workflow (Step 1 / Step 2, three node columns, Update Node Settings, INI settings files, CSV recording) is the same. nemi Studio adds live FFT, attitude and rotation-angle analysis, Big Plot, Common Plot, Node Zero and Scaling, Sound Output, resampling and nemi G+ M256 support. wirelessSensorPlot is no longer maintained.
Receiver and Radio Network¶
Can nemi Studio be used without a nemi Connect?
Yes, with a nemi EdgeBase configured with logging_mode 6 and reachable via VPN or local network (TCP port 9413). For direct local radio reception on a PC, use a nemi Connect.
How many sensors can nemi Studio handle? Up to 8 sensor nodes per radio channel (node count 1, 3 or 8). nemi Studio has three node columns; in 8-node mode, each column can show any of the eight nodes. For more than 8 sensors, use additional radio channels, each with its own receiver.
Why do I only see three columns if the network has eight nodes? The columns are display and configuration views, not the node limit. Use the node selection at the top left of each column to choose which node is shown. All nodes are active and recorded at the same time; the Node Status table shows all eight.
Can I mix nemi G+, nemi G+ ultra and nemi DAQ nodes on the same radio channel? Yes. Each node can be a different type. nemi Studio detects the type from the first data message and switches the column automatically.
Which radio data rate should I choose? Choose the slowest radio data rate that still supports the required sampling rates of all nodes: slower modes give more range but less channel capacity. nemi Studio pre-selects suitable default sampling rates. See nemi Link 2400 for details.
Can two receivers listen on the same radio channel at the same time? No. Two active receivers on the same channel cause collisions and data loss. Assign different channels to different sensor groups.
How do I change the radio channel during a measurement? After Start Receiver, use Change Channel; the change takes effect immediately. Data rate and number of nodes can only be changed after Disconnect COM-Port.
Configuration¶
Do I need to click Update Node Settings after changing a parameter? Yes. Red text on Update Node Settings means there are unsent changes. A node may take up to approx. one minute to apply new settings.
Are settings stored on the sensor node?
Only the radio parameters (channel, data rate, node number, node count), written with nemi Sensor Config via USB. Runtime settings are sent by nemi Studio; save them with Save Settings to File (INI). The last session is also stored automatically in last_session_settings.ini.
What happens if I load a settings file for the wrong sensor type? When data of the actual node arrives, nemi Studio corrects the module type automatically and sends the stored settings for that type.
Why are the ranges 16 … 256 g greyed out? They are only enabled when the node reports the matching hardware (nemi G+ M256 or High-G IMU) in its start-up messages. Switch the node off and on again while nemi Studio is connected.
How do I zero a strain gauge bridge or an offset? Use Node Zero and Scaling: Zero averages the signal for 3 s and sets the offset so that the current value becomes zero; you can also enter a scale factor per channel.
Analysis¶
Can nemi Studio show an FFT? Yes. Live Analysis → FFT shows the amplitude spectrum with selectable window function, window length and overlap. For detailed post-processing, use the recorded CSV files in MATLAB, GNU Octave or Python.
Can nemi Studio measure roll, pitch and yaw? Yes, in Live Analysis → Attitude via sensor fusion. Roll and pitch are available immediately; yaw requires the automatic compass calibration (rotate the sensor once, tilt it). Avoid magnetic mounting when yaw is needed.
Can I measure the rotation angle and speed of a shaft? Yes, in Live Analysis → Rotation Angle. Use a high IMU sampling rate and a gyroscope range matching the speed (4,000 °/s ≈ 666 rpm); at least approx. 8 samples per revolution are required. Vertical shafts are not supported.
Is nemi Studio suitable for precise tilt measurement? The Attitude tab gives roll and pitch for orientation purposes. For high-resolution static tilt measurement, record the primary accelerometer data (e.g. nemi G+ M08 at ±2 g, 125 Hz) and evaluate it with low-pass filtering, or use the Tilt Measurement view of the nemi Studio Web App.
Data and Files¶
Which files does nemi Studio record? CSV files per node for IMU data, vibration (nemi G+ / G+ ultra) or telemetry (nemi DAQ) and digital inputs, plus optional timing, debug and resampled files.
Which timestamp should I use for analysis?
Use Timecounter (s) for analysis within one recording. Use Timestamp (s) (UNIX time) to align data with other systems or several receivers.
What CSV import settings should I use?
Semicolon (;) as separator; the first line contains the column names.
Can I get all nodes in one file? Yes. Enable Resample Data and choose an output rate; after the recording all files are combined into one CSV file on a common time grid.
Is nemi Studio suitable for long-term unattended measurements? Only as long as the PC keeps running. For autonomous long-term monitoring use nemi EdgeBase or nemi Log; nemi Studio remains useful for commissioning and spot checks.
Time Synchronisation¶
Are the nodes synchronised with each other? Yes. Within one nemi Link 2400 network, all sensor clocks are synchronised to the receiver. Node Details shows the synchronisation accuracy of the last 100 synchronisation events.
Can several receivers record synchronously? Yes. Use several receivers on one computer, or several computers synchronised to the same NTP server. nemi Studio relates each receiver's time to the PC clock.
Support¶
What should I send to i4M when reporting a problem?
A recording with Log Debug only or the _debug.csv file of the measurement, and the console content copied with Copy Console / Copy to Clipboard.