nemi DAQ nano¶

Description¶
nemi DAQ nano is a small, wireless OEM data acquisition module in the form of a circuit board for highly precise, wireless acquisition of analog and digital signals. As a wireless strain gauge amplifier and telemetry module in OEM format, it offers versatile connection options for external sensors — for example to measure or monitor strain, forces, moments, distances (e.g. cracks) or temperatures. It combines two external analog channels with a 24-bit ADC, two digital inputs, an integrated 9-DoF IMU (accelerometer, gyroscope, magnetometer) and an internal temperature sensor — all on a board of 30 × 30 mm and less than 10 g including connectors.
Thanks to the integrated IMU, the module also captures its spatial orientation, which can save additional rotation-angle sensors, particularly in rotating applications. Because of its compact and lightweight design, nemi DAQ nano can be integrated into the smallest installation spaces and turns components and machines into smart products.
All measurement data is transmitted live over i4M's proprietary nemi Link 2400 radio technology in the 2.4 GHz band to a compatible receiver — nemi Connect (USB receiver for PC or Gantner Q.Station), nemi Log (compact SD-logger with cellular option), or nemi EdgeBase (autonomous field gateway with 4G, edge computing, and local SSD storage). All three receivers share the same star-topology radio architecture: one receiver per radio channel, up to eight sensor nodes per channel. The receiver handles data access, storage, downstream integration and time synchronization.
Power is supplied flexibly via a rechargeable Li-ion battery (with on-board charger), a non-rechargeable lithium battery, a power bank, the wide-range DC input (5–30 V) or Micro USB. An onboard 64 MHz 32-bit microcontroller supports edge computing so that pre-processed "smart data" can be transmitted instead of raw samples. Edge computing can be realized upon request in bilateral development projects.
The bare PCB has no housing protection rating. For permanent installations, nemi Struct offers a robust housing solution. The larger, housed variant with three analog and up to six digital inputs is the nemi DAQ.
This page describes hardware rev. 2.1 and later. For hardware rev. 1.x see Product Changes.
Typical Applications¶
- Torsion / torque and load measurements on rotating or moving components — e.g. in drive trains. The low mass and wireless operation eliminate slip rings and cable routing through rotating assemblies.
- Strain, force and displacement monitoring — full-bridge strain gauges, potentiometer sensors and 0–10 V transducers.
- Smart-product integration — embedding wireless sensing into components such as press-fit sensors, conveyor elements, machine parts or amusement ride structures.
- Structural health monitoring (SHM) — multi-node networks on bridges, machinery frames or buildings, with long-term operation from primary cells; housed in nemi Struct.
- Temperature measurement — Pt100 / Pt1000 resistance thermometers, thermocouples or thermistors.
- Vibration, inclination and rotation rate — using the integrated IMU without any external sensors.
Technical Specifications¶
General¶
| Parameter | Value | Unit |
|---|---|---|
| Dimensions (with connectors) | 32 × 32 × 12 | mm |
| Dimensions (without connectors) | 30 × 30 × 6 | mm |
| Weight (with connectors) | < 10 | g |
| Weight (without connectors) | < 7 | g |
| Operating temperature range | −25 to 85 | °C |
| Housing protection rating | none (bare PCB) | — |
| Onboard MCU, usable for edge computing | 64 MHz 32-bit MCU, 1 MB Flash, 256 KB RAM, various hardware crypto features | — |

ESD-sensitive device
nemi DAQ nano has limited built-in ESD protection. Damage may occur on devices subjected to high-energy electrostatic discharge. Take proper ESD precautions (wrist strap, ESD mat) when handling the board and connecting cables to avoid performance degradation or loss of functionality.
Power Supply¶
| Parameter | Value | Unit |
|---|---|---|
| Supply voltage on BAT pin — rechargeable battery | 3.3–4.2 | V |
| Supply voltage on BAT pin — non-rechargeable battery, power bank, other DC source | 3.3–5 | V |
| Supply voltage on DCIN pin | 5–30 | V |
| Micro USB (charging and power supply) | 5 | V |
| Charging cut-off voltage (rechargeable battery) | 4.2 | V |
| Charging current (rechargeable battery) | 120 | mA |
| Typical average consumption (see Power Consumption and Battery Life) | 3.64 mA at 3.3 V (≈ 12 mW) | — |
External Analog Inputs¶
| Parameter | Value | Unit |
|---|---|---|
| Number of analog channels | 2 | — |
| Connectable sensor types | Reference-voltage sensors (Wheatstone / strain gauge full bridges, potentiometer sensors, thermocouples e.g. Type K, thermistors); reference-current sensors (Pt100 / Pt1000); analog sensors with 0–10 V output | — |
| Max. sampling rate (with active channels) | 6,400 (1 channel); 1,944 (2 channels) | Hz |
| Signal bandwidth (−3 dB) | sampling rate × 0.23 | Hz |
| Sampling rate stability (full temperature range) | ± 50 | ppm |
| Signal resolution | 24 | bit |
| Analog measurement error | << 0.1 (with automatic self-calibration) | % |
| Selectable signal gains | 1 / 4 / 8 / 16 / 32 / 64 / 128 | — |
| Reference voltage / supply voltage | 2.5 | V |
| Reference current (Pt100 / Pt1000) | 500 | µA |
External Digital Inputs¶
| Parameter | Value | Unit |
|---|---|---|
| Number of digital channels | 2 | — |
| Digital input signal level (high) | 3.3–24 | V |
| Sampling rate | 2,000 | Hz |
| Sensor switching voltage, high level | 3.3–24 (selectable in software) | V |
| Sensor switching voltage, low level | < 2 | V |
| Compatible sensors | Switching displacement, distance, level or rotation-angle sensors; TTL / NPN sensors | — |
Integrated 9-DoF IMU (ACC / GYR / MAG)¶
| Parameter | Value | Unit |
|---|---|---|
| Sensor type | 9-DoF MEMS IMU: 3-axis accelerometer, 3-axis gyroscope, 3-axis magnetometer | — |
| ACC / GYR sampling rates | 3,840 / 1,920 / 960 / 480 / 240 / 120 / 60 | Hz |
| MAG sampling rate | always 1/8 of the ACC / GYR rate | — |
| Signal resolution | 16 | bit |
| ACC ranges | ± 256 / 128 / 64 / 32 / 16 / 8 / 4 / 2 | g |
| GYR ranges | ± 4,000 / 2,000 / 1,000 / 500 / 250 / 125 | °/s |
| MAG ranges | ± 16 / 12 / 8 / 4 | Gauss |
Internal Temperature Sensor¶
| Parameter | Value | Unit |
|---|---|---|
| Measuring range | −20 to 60 | °C |
Radio (nemi Link 2400)¶
| Parameter | Value |
|---|---|
| Frequency band | 2.4 GHz |
| Topology | Star (one receiver per channel) |
| Max. sensor nodes per channel | 8 |
| Radio modes | 2 Mbit/s / 1 Mbit/s / 500 kbit/s / 125 kbit/s |
| Range | up to 350 m (see Range Considerations) |
| Antenna connection | u.FL connector on the board (for an external antenna) |
| Compatible receivers | nemi Connect, nemi Log (+ cellular), nemi EdgeBase |
First Commissioning¶
Required Hardware and Software¶
| Item | Purpose |
|---|---|
| nemi DAQ nano with the ordered configuration (K option, power option) | Sensor node |
| 2.4 GHz antenna with u.FL connector — only antennas approved by i4M (not included as standard, available on request) | Radio link |
| Power source | Rechargeable / non-rechargeable battery, power bank, DC supply 5–30 V or Micro USB |
| nemi Connect / nemi EdgeBase / nemi Log | Receiver |
| Micro-USB cable (data-capable, not charge-only) | USB configuration and charging |
| nemi Sensor Config v2.1 or later | Radio and node configuration via USB |
| nemi Studio v1.1 or later or nemi Studio Web App | Wireless measurement settings, live display and CSV recording |
Antenna required — use only approved antennas
Before operating the radio, an antenna must be connected to the u.FL connector of the board. No antenna is included as standard; i4M supplies approved antennas on request. Only antennas approved by i4M may be used — please ask i4M for approved antennas.
Download links for the software tools are provided to customers (the software is not publicly available).
Step 1 — Identify Hardware Revision and Configuration¶
- Check the version printed on the board (v2.x). This page applies to hardware rev. 2.1 and later; for rev. 1.x see Product Changes.
- Check the ordered analog configuration (K01–K04, see Mounting & Connections). The input configuration and the power option are defined before delivery; subsequent modification is only possible by i4M technologies.
- Check the solder jumpers on the bottom side of the board against your power source (rechargeable or non-rechargeable battery, see Power Supply Options).
Step 2 — Connect Antenna, Sensors and Power¶
- Attach the antenna to the u.FL connector — only antennas approved by i4M (not included as standard; available from i4M on request). If the board is mounted inside a metal enclosure, route an external antenna to the outside.
- Wire the external sensors according to the K option (see Pin Assignment). Observe ESD precautions.
- Connect the power source to the 4-pin power terminal (BAT or DCIN) with correct polarity and set the slide switch accordingly (top = BAT, bottom = DCIN). Alternatively, supply the board via Micro USB.
The LED shows the restart sequence (white → green → red → blue) when the board starts. When a rechargeable battery is used, the solid red LED indicates charging via USB.
Step 3 — Configure the Radio Parameters via USB¶
Radio parameters (channel, radio mode, node number, maximum nodes) are persistent settings written with nemi Sensor Config over USB. The procedure is the same as for all nemi Link 2400 sensor nodes (see nemi G+ for screenshots):
- Launch nemi Sensor Config and connect the board to the PC with a data-capable Micro-USB cable.
- After the restart sequence the LED shows solid green + flashing yellow (USB detected; waiting for COM port).
- nemi Sensor Config shows the new nemi sensor node and selects its COM port automatically. Click Connect COM-Port within approximately 10 seconds; otherwise the node enters radio transmission mode. The LED changes to solid green + flashing blue.
- Enter the radio parameters (see table below) and click Set nemi Link Parameters.
- The node prints its new radio parameters in the console; after 10 s it starts radio transmission with the new parameters.
Recommended first-test radio settings:
| Parameter | Recommended value | Notes |
|---|---|---|
| nemi Link 2400 Mode | 2 Mbit/s | Highest throughput; adequate range for bench tests |
| nemi Link 2400 Node Count | 1 or 3 | Increase to 8 nodes if needed, but the maximum sampling rate per node is reduced |
| nemi Link 2400 Channel | 0 (other preferred channels are 25, 50, 75) | Default and a good choice for most installations |
| nemi Link 2400 Node Number | 1 | Unique per node on the same channel |
If the nemi DAQ nano is ordered without USB connector (e.g. due to height restrictions), the radio parameters are set by i4M according to the order (see Ordering Options).
Step 4 — Set Up the Receiver and Verify Data¶
With nemi Connect (or a nemi Log used as USB receiver):
- Launch nemi Studio or open the nemi Studio Web App in Chrome or Edge.
- Plug the nemi Connect into a USB port. The nemi Connect is detected and its COM port is selected automatically.
- Click Connect COM-Port, set the radio settings matching the node, select the module type nemi DAQ nano and click Start Receiver.
- Power the nemi DAQ nano. Within a few seconds live data should appear. Check the signal strength and latency indicators.
- Set the analog sampling rate (AF ODR), the gain per channel, the IMU rate and ranges and the digital input level (GPIO) wirelessly and click Update Node Settings. A node may take up to approx. one minute to apply new settings.
- Select the display unit of the analog channels (Y Axis Unit: mV/V, V, °C for Pt100 / Pt1000, or 0–10 V), apply a known load or excitation and start a short CSV recording to verify the data before final installation.
With nemi Log or nemi EdgeBase: these receivers work standalone and are configured via CSV (nemi Log) or YAML files (nemi EdgeBase). The nemi DAQ nano pairs automatically when channel and radio mode match. Use EdgeBase logging_mode 6 to check live data with nemi Studio or the nemi Studio Web App over the network.
LED Status Codes¶
| LED pattern | Meaning |
|---|---|
| Rapid flash: white → green → red → blue | Sensor node restart |
| Solid red | Battery charging via USB or wide-range voltage input |
| Red no longer continuously lit | Battery fully charged |
| Solid green + flashing yellow | USB data connection detected; waiting for COM port to open |
| Solid green + flashing blue | COM port open; waiting for RF parameters or settings |
| Short red flashes (~2 s) | Actively sampling and transmitting measurement data |
| Short green flashes every 10 s (~1 s) | Sleep mode |
| Short blue flashes (~1 s) | Time synchronization with wireless network |
| Green LED on temporarily or continuously (in wireless mode) | Internal ring buffer 80 % full — data cannot be transmitted fast enough; risk of data loss |
Range Considerations¶
nemi Link 2400 operates at 2.4 GHz. Typical ranges by radio mode:
| Radio mode | Typical range (in buildings – in urban environment) |
|---|---|
| 2 Mbit/s | 20 – 50 m |
| 1 Mbit/s | 50 – 125 m |
| 500 kbit/s | 90 – 225 m |
| 125 kbit/s | 140 – 350 m |
Range depends strongly on the antenna and its placement. Metal enclosures block 2.4 GHz radio — route the antenna to the outside of any metal housing and always validate the range on site.
Mounting & Connections¶
Physical Mounting¶
The nemi DAQ nano is a bare PCB with two mounting holes (24 mm spacing in both directions, see dimension drawing). It is designed for integration into customer-provided enclosures or into nemi Struct.
- The board has no housing protection rating. For outdoor, humid or mechanically exposed installations, use a sealed enclosure.
- Metal enclosures block 2.4 GHz radio — route an external antenna to the outside.
- When the IMU is used for vibration or inclination measurements, mount the board rigidly to the measured structure.
Configuration Options¶
nemi DAQ nano has 2 analog and 2 digital inputs. The analog inputs are configured before delivery:
- Up to 2 reference-voltage sensors (e.g. Wheatstone / strain gauge full bridges, potentiometer sensors, thermocouples e.g. Type K, thermistors)
- Up to 2 reference-current sensors (e.g. Pt100 / Pt1000 resistance thermometers)
- Up to 2 sensors with 0–10 V output (with attenuator)
The digital inputs accept up to 2 switching digital sensors (e.g. switching displacement, distance, level or rotation-angle sensors) or TTL / NPN sensors.

| K option | Channel 1 | Channel 2 | Digital inputs |
|---|---|---|---|
| K01 | Reference voltage | Reference voltage | 2 |
| K02 | Reference voltage | 0–10 V | 2 |
| K03 | 0–10 V | 0–10 V | 2 |
| K04 | Reference current | Reference current | 2 |
A combination of reference-voltage sensors and reference-current sensors on one nemi DAQ nano is not possible.
Pin Assignment¶
All connections of the nemi DAQ nano 2.x are located on the top side of the board:
- 4-pin power terminal (GND, BAT, RES, DCIN) for the battery, a power bank or an external DC supply of 5–30 V. Do not connect anything to the RES pin (reserved for internal functions).
- 8-pin analog terminal with the two analog input channels (GND, CH+, CH−, V_ref per channel).
- Two digital inputs (DIN 1, DIN 2, GND). No terminal is fitted for the digital inputs; the wires are soldered to the solder pads (DIN 1, DIN 2 and 2 × GND).
- Slide switch to choose the power source (BAT pin or DCIN pin).
- Micro USB connector for charging and power supply.

CH+ / CH− = measuring input; V_ref = reference voltage (2.5 V); GND = ground; BAT = battery; RES = reserved; DCIN = DC input 5–30 V; DIN = digital input.
Pin assignment of the analog inputs per configuration:
| Terminal pin | K01 | K02 | K03 | K04 |
|---|---|---|---|---|
| GND | GND | GND | GND (sensor GND) | GND (white) |
| CH1+ | CH1+ | CH1+ | 0–10 V output | CH1+ (red) |
| CH1− | CH1− | CH1− | bridged to GND | CH1− (white) |
| V_ref | V_ref (2.5 V) | V_ref (2.5 V) | n.c. | n.c. |
| GND | GND | GND (sensor GND) | GND (sensor GND) | GND (white) |
| CH2+ | CH2+ | 0–10 V output | 0–10 V output | CH2+ (red) |
| CH2− | CH2− | bridged to GND | bridged to GND | CH2− (white) |
| V_ref | V_ref (2.5 V) | n.c. | n.c. | n.c. |
n.c. = not connected. K04: wire colors of a 3-wire Pt100 / Pt1000 sensor.
For 0–10 V sensors, always connect the sensor GND to the GND pin of the channel to reduce noise. As the sensor output is referenced to GND, CH− is bridged to GND.
Example Wiring Diagrams¶
| K01: strain gauge full bridge + potentiometer | K02: reference voltage + 0–10 V |
|---|---|
![]() |
![]() |
| K03: 2 × 0–10 V | K04: 2 × Pt100 / Pt1000 (3-wire) |
![]() |
![]() |
V_m = measuring voltage; n.c. = not connected.
Digital inputs (logic high 3.3–24 V):

Power Supply Options¶
By default, nemi DAQ nano 2.x is powered by a rechargeable Li-ion battery. Optionally — set with solder jumpers — a single non-rechargeable lithium battery (Li/SOCl₂) or 3 alkaline batteries in series can be used. The 5–30 V DC input is available with every power option. The Micro-USB connector can also be used as power supply (5 V).
Check solder jumpers and slide switch before connecting power
The power supply is configured via solder jumpers on the bottom side of the board. Before connecting the power source, make sure the correct solder jumpers are set. Rechargeable Li-ion batteries must have their own protection circuit.
Selecting the power source with the slide switch:
| Slide switch | Power source | Examples |
|---|---|---|
| Top position | BAT pin | Rechargeable Li-ion battery (with protection circuit), non-rechargeable lithium battery or other DC supply 3.3–5 V |
| Bottom position | DCIN pin | 24 V DC, 2–8 non-rechargeable lithium batteries in series, 12 V lead-acid battery |
In the other switch position the board is either off, or the switch can be used to change between the two power sources if both are connected.
| Powered by BAT pin | Powered by DCIN pin |
|---|---|
![]() |
![]() |
Solder jumper configuration (bottom side of the board, can be adapted by the customer):
| Power option | Solder jumper function |
|---|---|
| Rechargeable battery (default) | The BAT pin is connected to the charger IC so that the battery can be charged with the internal charging circuit (via USB). The battery voltage is measured. Charge characteristic: constant current – constant voltage, 4.2 V, 120 mA. |
| Non-rechargeable battery | The BAT pin is not connected to the charger IC (prevents accidental charging of a non-rechargeable battery). The battery voltage is measured. Input range: 3.3–5 V. |

Antenna¶
The nemi DAQ nano has a u.FL connector for connecting an external 2.4 GHz antenna. No antenna is included as standard; i4M supplies approved antennas on request. Only antennas approved by i4M may be used — please ask i4M for approved antennas. For installations in metal enclosures, use an external 2.4 GHz, 50 Ω antenna routed to the outside (e.g. via a u.FL → RP-SMA pigtail and a bulkhead adapter).
Detailed Configuration¶
Configuration is split into three layers:
- Hardware configuration — K option and power option (solder jumpers), defined before delivery.
- Persistent radio settings — written via
nemi Sensor Configover USB (channel, radio mode, node number, max nodes). - Runtime measurement settings — set wirelessly via nemi Studio, nemi Studio Web App, nemi EdgeBase YAML files or nemi Log configuration CSVs (active channels, sampling rates, gains, IMU ranges, digital input levels, sleep mode).
Configuration Tools Reference¶
| Tool | Connection | Configures |
|---|---|---|
| nemi Sensor Config v2.1+ | USB (Micro USB) | Radio channel, radio mode, node number, max nodes per channel |
| nemi Studio | via nemi Connect or nemi EdgeBase | Mode (sleep / continuous / intermittent), AF ODR, gain per channel, IMU ODR and ranges, GPIO level, zero and scaling |
| nemi Studio Web App | via nemi Connect or nemi EdgeBase | Radio network, sampling rates, measuring ranges, live display and CSV recording |
| nemi Link 2400 Radio Channel Calculator | Online tool (replaces the former Excel configuration tool) | Validates that sampling rates fit within channel bandwidth for given radio mode and node count |
| nemi EdgeBase YAML config files | VPN / FTP | Sampling rates, gains, logging intervals, MQTT settings, sleep/wake schedules |
| nemi Log CSV config files | SD card / FTP | Sampling rates, gains, logging intervals, sleep/wake schedules |
Analog Channels: Sampling Rate¶
The live radio-transmitted sampling rate applies to all active channels and depends on the number of active channels. For the highest sampling rate (6,400 Hz), activate only one channel.
| ODR setting | Mode | 1 channel (Hz) | 2 channels (Hz) |
|---|---|---|---|
| 0 | Off | AD off | AD off |
| 1 | High Res | — | 1,944 |
| 2 | High Res | 6,400 | 1,074 |
| 3 | High Res | 4,800 | 742 |
| 4 | High Res | 3,200 | 567 |
| 5 | High Res | 2,400 | 459 |
| 6 | High Res | 1,600 | 385 |
| 7 | High Res | 800 | 332 |
| 8 | High Res | 400 | 291 |
| 9 | High Res | 200 | 235 |
| 10 | High Res | 100 | 196 |
| 11 | Ultra Low Power | 100 | 100 |
| 12 | Ultra Low Power | 75 | 75 |
| 13 | Ultra Low Power | 50 | 50 |
| 14 | Ultra Low Power | 25 | 25 |
| 15 | Ultra Low Power | 1 | 1 |
The −3 dB signal bandwidth equals the sampling rate multiplied by 0.23: 1,472 Hz at 6,400 Hz, 23 Hz at 100 Hz.
- High Res (ODR 1–10): continuous bridge excitation; suitable for all cable lengths; higher power consumption.
- Ultra Low Power (ODR 11–15): pulsed bridge excitation to save energy. Pulsed excitation can cause capacitive spike artefacts with long sensor cables — use High Res mode if spikes are observed.
Analog Channels: Gain and Measuring Range¶
The signal gain is configured separately for each channel. With gain 1, the measuring range of reference-voltage sensors is 1,000 mV/V (0–2.5 V at 2.5 V supply voltage); with a Pt100 reference-current sensor the absolute measuring range is 0–249 mV. A 0–10 V configuration extends the input range to 0–10 V.
| Gain setting | Amplification | Reference voltage (mV/V) | Reference voltage, absolute (mV) | Reference current, absolute (mV) | Pt100 range (°C) | Pt1000 range (°C) | 0–10 V input (V) |
|---|---|---|---|---|---|---|---|
| 0 | Channel off | — | — | — | — | — | — |
| 1 | 1 | 1,000.00 | 2,500.00 | 249 | > 850 | 400 | 10.00 |
| 2 | 4 | 250.00 | 625.00 | 62.25 | > 850 | — | 2.50 |
| 3 | 8 | 125.00 | 312.50 | 31.13 | 590 | — | 1.25 |
| 4 | 16 | 62.50 | 156.25 | 15.56 | 145 | — | 0.63 |
| 5 | 32 | 31.25 | 78.13 | 7.78 | −60 | — | 0.31 |
| 6 | 64 | 15.63 | 39.06 | 3.89 | −155 | — | 0.16 |
| 7 | 128 | 7.81 | 19.53 | 1.95 | −200 | — | 0.08 |
For small bridge signals (typical strain gauge measurement), the highest gain (128×) is the usual starting point. For 0–10 V transducers, gain 1 covers the full range. For potentiometers that show unstable behaviour, gain 4 is a recommended fallback.
IMU Configuration¶
ACC and GYR share the selected IMU rate; the MAG always runs at 1/8 of the ACC / GYR rate.
| IMU ODR setting | ACC / GYR (Hz) | MAG (Hz) |
|---|---|---|
| 0 | IMU off | IMU off |
| 1 | 60 | 7 |
| 2 | 120 | 15 |
| 3 | 240 | 30 |
| 4 | 480 | 60 |
| 5 | 960 | 120 |
| 6 | 1,920 | 240 |
| 7 | 3,840 | 480 |
| Sensor | Selectable ranges | Selection guidance |
|---|---|---|
| ACC | ±2 / 4 / 8 / 16 / 32 / 64 / 128 / 256 g | ±2 g for inclination with small dynamic accelerations; higher ranges for rotating parts with large centrifugal acceleration or for shocks |
| GYR | ±125 / 250 / 500 / 1,000 / 2,000 / 4,000 °/s | Select above the expected peak rotation rate (4,000 °/s ≈ 666 rpm) |
| MAG | ±4 / 8 / 12 / 16 Gauss | Enable only when heading information is required |
In nemi Studio, the extended ACC ranges (32–256 g) are enabled automatically as soon as the node reports the high-g IMU when it starts.
Digital Inputs¶
The two digital inputs are sampled at a fixed 2,000 Hz. The high level is selected in software (nemi Studio setting GPIO: off, 5 V, 12 V, 24 V); the low level must be below 2 V. The 5 V setting is also used for 3.3 V signals, as the high level threshold is the same. Verify that 2,000 Hz is sufficient for the maximum event frequency of your application.
Radio Mode, Channel and Node Configuration¶
| Parameter | Range | Notes |
|---|---|---|
| Radio mode | 2 Mbit/s, 1 Mbit/s (HS) / 500 kbit/s, 125 kbit/s (XR) | Must match the receiver |
| Radio channel | 0–79 | Receiver and all nodes on the same network must match; preferred channels 0, 25, 50, 75 |
| Max. nodes | 1, 3 or 8 | Defines bandwidth allocation; must match the receiver |
| Node number | 1 … max. nodes | Unique per radio channel |
Verify every planned combination of nodes, sampling rates and radio mode with the nemi Link 2400 Radio Channel Calculator.
Sleep Mode¶
The nemi DAQ nano supports a low-power sleep mode and intermittent operation (measure for a configured time, then sleep) to extend battery life:
- The sleep interval is configured in 30-second steps; the maximum uninterrupted sleep is 65,535 s (~18.2 h).
- During sleep, the LED shows short green flashes every 10 seconds.
- The node wakes, checks for transmission and returns to sleep automatically.
0–10 V Conversion Factor¶
For 0–10 V inputs (K02, K03), the physical voltage is recovered from the mV/V output by:
nemi Studio applies this automatically when the display unit 0 - 10 V is selected. An early prototype used a different factor (10.09091); if results are approximately 14 % off, verify the conversion factor in use.
Power Consumption and Battery Life¶
With the following example settings, nemi DAQ nano consumes an average of 3.64 mA at 3.3 V (12 mW) on the BAT pin: 50 Hz analog ODR with two Pt100 / Pt1000 sensors, 120 Hz IMU ODR with ACC, GYR and MAG active, DIN off, radio mode 2 Mbit/s.
| Power source | Connected to | Nominal capacity (Wh) | Efficiency at nominal voltage | Effective capacity (Wh) | Projected runtime (days) |
|---|---|---|---|---|---|
| 1 × Saft LS 14250 | GND and BAT | 4.3 | ~89 % (3.6 V) | 3.8 | 13 |
| 1 × Saft LS 33600 | GND and BAT | 61.2 | ~89 % (3.6 V) | 54.5 | 189 |
| 2 × Saft LS 33600 in series | GND and DCIN | 122.4 | ~85 % (7.2 V) | 104 | 361 |
| 4 × Saft LS 33600 in series | GND and DCIN | 244.8 | ~81 % (14.4 V) | 198 | 688 |
The runtime is calculated from the nominal battery capacity; the usable capacity depends on temperature, discharge rate and other environmental factors. Strain gauge bridges can dominate the consumption: two 350 Ω bridges at 2.5 V draw approximately 14 mA. Use higher-resistance bridges (e.g. 1,000 Ω) where the application permits.
For Saft Li-SOCl₂ batteries, the voltage can drop below 3.3 V at temperatures below 0 °C. A single-battery configuration is therefore not recommended for applications below 0 °C.
Power source examples:
| Type | Example | Nominal voltage (V) | Capacity (Wh) | Dimensions (mm) |
|---|---|---|---|---|
| Rechargeable (LiPo) | Renata ICP641620PA | 3.7 | 0.6 | 7 × 16 × 22 |
| Rechargeable | Panasonic NCR18650B | 3.7 | 12.6 | Ø 18 × 69 |
| Non-rechargeable (Li-SOCl₂) | Saft LS14250 | 3.6 | 4.3 | Ø 14 × 25 |
| Non-rechargeable (Li-SOCl₂) | Saft LS33600 | 3.6 | 61 | Ø 33 × 61 |
| Non-rechargeable (lithium primary) | Varta 9 V battery | 9 | 10.8 | 26 × 49 × 16 |
| Non-rechargeable (alkaline primary) | Varta Industrial Pro 4020 D Mono (3 in series) | 4.5 | 76.5 | 3 × Ø 34.2 × 61.5 |
Time Synchronization¶
Within one nemi Link 2400 wireless network, the nemi DAQ nano clock is synchronized to the receiver module with an accuracy of ±10 µs with a stable radio connection. With a poor radio connection, some synchronization events have to be discarded because their transit time is too long, and the accuracy decreases. Short blue LED flashes indicate a resynchronization. The receiver handles its own downstream synchronization (nemi EdgeBase: NTP or GNSS; nemi Log: RTC or GNSS; nemi Connect with nemi Studio: host computer clock), documented with the respective receiver product.
Integration¶
nemi DAQ nano communicates only via nemi Link 2400 radio to a receiver. All external integration happens at the receiver level and is identical to the nemi DAQ:
| Receiver | Best for | Data access |
|---|---|---|
| nemi Connect | First tests and local measurements with nemi Studio or nemi Studio Web App, custom USB integration with the nemi Connect Python API, Gantner Q.Station | USB virtual COM port |
| nemi Log | Compact local medium-term standalone logging, optional cellular upload | nemi Log .bin files with Python decoder |
| nemi EdgeBase | Unattended long-term monitoring, MQTT streaming, raw file storage, edge computing | .bin.bz2 files via Ethernet, LTE, OpenVPN, FTP; MQTT(S) streams |
Integration Architecture¶
nemi DAQ nano
│ nemi Link 2400 (2.4 GHz wireless)
▼
Receiver module
├── nemi Connect → USB → PC (nemi Studio / nemi Studio Web App / Python API / Q.Station)
├── nemi Log → SD card / LTE Cat M or NB-IoT with cellular option
└── nemi EdgeBase → Ethernet / LTE / OpenVPN
├── MQTT broker (aggregated values) → Database → Dashboard
└── FTP server (raw compressed binary files) → Python decoder → pandas DataFrames
MQTT Streams (via nemi EdgeBase)¶
Topic structure: <mqtt_root_topic>/<sensor>/<stream>; each message carries all channels of one stream as min / max / mean values over the send interval. See nemi DAQ — MQTT Streaming and nemi EdgeBase for configuration and payload format.
| Stream | Content | Channels used by the nemi DAQ nano |
|---|---|---|
STREAM_0 |
Analog channels | TIMECOUNTER, CH_1, CH_2, VOLT, RSSI |
STREAM_1 |
IMU | TIMECOUNTER, ACC_X/Y/Z (g), GYR_X/Y/Z (°/s), MAG_X/Y/Z (Gauss), TEMP, VOLT, RSSI |
STREAM_2 |
Digital inputs | TIMECOUNTER, DIN_1, DIN_2 |
The stream definition is shared with the nemi DAQ (CH_1 … CH_4, DIN_1 … DIN_6); channels not present on the DAQ nano carry no meaningful data (CH_4 is only used by customer-specific four-channel modules). Iterate the keys present in the payload.
Binary Data Format¶
Data recorded on the EdgeBase is stored as compressed binary files (.bin.bz2, on the nemi Log as .bin); they contain the same COBS-encoded binary messages as the USB stream of the nemi Connect. Analog channels, IMU and digital inputs are transmitted in separate data messages, each with the synchronized timestamp (1 µs resolution) of its last sample. i4M provides Python decoders for the efficient binary files of nemi EdgeBase (.bin.bz2) and nemi Log (.bin). They decode the files into pandas DataFrames with timestamps and values in physical units, which can then be saved as CSV, Parquet etc. or processed directly in a Python environment. The complete protocol documentation is provided to customers on request; i4M discloses the protocol to its customers.
Gantner Instruments Q.Station¶
The nemi DAQ nano is compatible with the Gantner Instruments Q.Station via nemi Connect and the GI nemione plugin. Configure it as a DAQ-type node with only the channels present on the DAQ nano active.
Troubleshooting¶
Start-up and Power¶
| Symptom | Likely cause | Resolution |
|---|---|---|
| No LED at all on power-up | Wrong slide switch position or incorrect solder jumpers | Check slide switch (top = BAT, bottom = DCIN) and solder jumpers for your power source |
| No LED at all on power-up | Depleted battery | Charge via Micro USB (rechargeable option) or replace the primary cell |
| Device starts and then shuts off after some time | Consumer power bank auto-shutoff at low current | Use an always-on power bank (e.g. Voltaic V75) or a direct battery / DC supply |
| USB disconnects during configuration | USB current transients (hubs, long cables) | Connect directly to a PC USB port with a short, high-quality cable; ensure the battery has some charge |
| USB connected but no COM port appears | Charge-only USB cable | Use a data-capable Micro-USB cable |
| Node works only while USB is connected | Battery, battery wiring or power path | Test with a known-good protected battery or a regulated DC supply on the intended input; check slide switch and solder jumpers |
| Repeated restarts under load | Supply voltage dips during radio transmission or bridge excitation | Measure the supply voltage under load; use a lower-impedance power source |
Radio and Connectivity¶
| Symptom | Likely cause | Resolution |
|---|---|---|
| Node transmits (red LED flashes) but does not appear in nemi Studio | Channel, radio mode or node count mismatch | Verify node settings with nemi Sensor Config; set the receiver to identical channel, mode and node count |
| Node does not appear | Node number conflict | Assign unique node numbers per channel |
| Node does not appear on nemi EdgeBase | Outdated EdgeBase RF controller firmware | Update the EdgeBase firmware (RF controller update requires physical access — contact i4M) |
| Green LED stays on in wireless mode | Ring buffer ≥ 80 % — data generated faster than transmitted | Reduce sampling rate or active channels; improve antenna placement; check RSSI |
| Blue flashes persist, no data | Node keeps resynchronizing; unstable receiver contact | Check receiver availability, antenna, distance and radio configuration |
Data Quality¶
| Symptom | Likely cause | Resolution |
|---|---|---|
| Periodic spikes to range limits on strain gauge channel | Ultra Low Power (pulsed excitation) mode + cable capacitance | Switch to High Res mode; alternatively apply a 3- or 5-point median filter in post-processing or in the EdgeBase MQTT path |
| 0–10 V values approx. 14 % off | Prototype conversion factor used | Use factor 8.647058824 |
| Analog channel shows constant value, excessive noise or wrong offset after cabling | ESD damage to the analog input | No field repair; return the board for inspection. Prevention: ESD precautions; connect sensor cables before powering |
| MQTT has gaps but EdgeBase raw files are complete | MQTT messages are not buffered during connectivity outages | Keep raw file logging enabled alongside MQTT |
| Custom decoder delivers implausible values | Frames parsed before COBS decoding | Split the stream at the 0 bytes and COBS-decode each message before parsing; or use the i4M Python decoder |
| Analog, IMU and digital streams have different timestamps | Each stream runs at its own native rate | Align by timestamps; resample only if the downstream system requires a common time grid |
Product Changes¶
Ordering Options¶
| Order code | Description |
|---|---|
| nemi DAQ nano Kxx | Analog configuration K01–K04; power option (rechargeable / non-rechargeable battery) to be specified |
| Version without USB connector | For height-restricted installations. Specify radio mode (2 Mbit/s, 1 Mbit/s, 500 kbit/s, 125 kbit/s), radio channel, maximum node count and node number with the order, as they cannot be set by the customer via USB |
Hardware Revisions¶
nemi DAQ nano is available in two hardware generations. The renderings and the table below help to identify your version. For hardware rev. 1.x please refer to the datasheet of the previous version (contact i4M if needed). If you need rev. 1.x due to size limitations, please contact i4M.

| Feature | Hardware rev. 1.x | Hardware rev. 2.x |
|---|---|---|
| Board size (without connectors) | 21 × 21 mm | 30 × 30 mm |
| Power supply | Battery with energy harvesting module (no longer available) | Battery, power bank or DC input 5–30 V; slide switch for the power source |
| Version printed on the board | V1.x | v2.x |
- v2.0: revised power section with DCIN 5–30 V, slide switch, solder-jumper power options and terminal-block connectors. Reference-voltage / reference-current mode selection depended on the firmware.
- v2.1: dedicated solder jumper for the reference-voltage / reference-current mode selection — no longer firmware-dependent.
Legacy rev. 1.x (summary): 15-pin connector (pins 1–4: power / battery — reversing polarity damages the board; pins 5–12: analog channels; pins 13–15: DIN 1, DIN 2, GND), Li-ion / LiPo charger up to 100 mA (charging cut-off 4.2 V, discharge cut-off 3.3 V). The energy harvesting (solar) option of rev. 1.x is no longer available.
Firmware Notes¶
| Issue | Affected hardware | Status |
|---|---|---|
| USB configuration tool | firmware < 10.0 | Use legacy wiredSensorConfig; firmware ≥ 10.0 requires nemi Sensor Config |
Software Tools Version Reference¶
| Tool | Version reference |
|---|---|
| nemi Sensor Config | 2.1 (as of March 2024) |
| nemi Studio | 1.1 Beta 2 (as of November 2026) |
| nemi Studio Web App | 2026.11.04.3 (as of November 2026) |
| wirelessSensorPlot | 7.4 (legacy, no longer maintained) |
FAQ¶
Applications¶
Can the nemi DAQ nano be used for strain gauge measurements? Yes. Configuration K01 supports two full-bridge strain gauges with 2.5 V excitation; K02 supports one bridge plus one 0–10 V sensor. For small bridge signals, gain 128 (7.81 mV/V full scale) is the typical starting point.
Can it be used on rotating machinery? Yes — this is one of the primary use cases. When using the IMU, mount the board as close to the rotation axis as possible; centrifugal acceleration adds a radial offset. The high-g IMU ranges up to ±256 g avoid saturation on fast-rotating parts.
Is it suitable for structural health monitoring? Yes. Several nodes can be distributed across a structure with one nemi EdgeBase as receiver, storing raw data locally and streaming reduced values via MQTT. With Saft LS33600 primary cells, runtimes of several months to more than a year are possible (see Power Consumption and Battery Life). For a robust housing, use nemi Struct.
Can it measure vibration? Yes, via the integrated IMU (up to 3,840 Hz ACC / GYR) or via external sensors on the analog inputs (up to 6,400 Hz on a single channel). The signal bandwidth is approximately sampling rate × 0.23.
Does it have GNSS? No. GNSS is a function of the nemi EdgeBase and nemi Log receivers.
Connectable Sensors¶
Can I mix strain gauges and Pt100 / Pt1000 on the same unit? No. Strain gauges require constant-voltage excitation (K01, K02); Pt100 / Pt1000 require constant-current excitation (K04). Use separate DAQ nano nodes. Thermistors are compatible with the reference-voltage configuration.
Can I use 0–10 V transducers? Yes, with a K02 or K03 unit. Connect the sensor GND to the channel's GND pin. Gain 1 covers the full 0–10 V range.
Can I use TTL or NPN switching sensors? Yes, via the two digital inputs (2,000 Hz sampling rate, low level < 2 V). The wires are soldered to the DIN solder pads.
Can I connect RS232 sensors directly? No. The nemi DAQ nano has analog and digital inputs only; serial sensors must be integrated at system level.
Can I connect geophones? Yes, if the geophone output voltage is within the analog input range of the selected gain.
Power Supply¶
Which power sources can I use? A rechargeable Li-ion battery with protection circuit (BAT pin, 3.3–4.2 V, default option), a non-rechargeable lithium battery or other DC source (BAT pin, 3.3–5 V, solder jumper option), 3 alkaline batteries in series, a DC supply or battery stack of 5–30 V on DCIN (e.g. 24 V DC, 12 V lead-acid battery, 2–8 lithium cells in series), or 5 V via Micro USB.
Can I power the node from 24 V DC? Yes, via the DCIN pin (5–30 V) with the slide switch in the bottom position.
Can I use a standard USB power bank? Consumer power banks often switch off when the current draw falls below their threshold, e.g. in sleep or low-power operation. Use always-on models (e.g. Voltaic V75) or a direct battery / DC supply for unattended deployments.
Can I use Li-SOCl₂ cells at low temperatures? Below 0 °C, the voltage of Li-SOCl₂ cells can drop below 3.3 V. A single-cell configuration on the BAT pin is not recommended below 0 °C; use several cells in series via DCIN instead.
Why do strain gauge bridges reduce the battery life? The bridge excitation current can dominate the total consumption: two 350 Ω full bridges at 2.5 V draw about 14 mA. Higher-resistance bridges (e.g. 1,000 Ω) greatly improve the runtime.
How long does the battery last? At the example configuration (≈ 12 mW): approx. 13 days with one Saft LS14250, 189 days with one LS33600, 361 days with two LS33600 in series and 688 days with four LS33600 in series. Actual runtime depends on active channels, sampling rate, radio mode, bridge resistance and temperature.
Radio and Integration¶
What radio range can I expect? Up to 350 m. Typical ranges are 20–50 m at 2 Mbit/s and 140–350 m at 125 kbit/s, depending strongly on antenna, environment and obstructions. Always test on site.
Can I use the board inside a metal enclosure? Yes, with an external antenna: connect a 2.4 GHz antenna to the u.FL connector and route it to the outside of the enclosure. Metal shields the radio signal.
Are all channels sampled simultaneously? No. Analog, IMU and digital streams have their own sampling rates and timestamps. The common receiver time base lets you align them, but the samples are not taken at identical instants.
Which software do I need to get started?
nemi Sensor Config (v2.1 or later) for the radio settings via USB, and nemi Studio or the nemi Studio Web App for wireless measurement settings, live display and CSV recording.
Can I decode binary files without i4M software?
Yes. Decompress the .bin.bz2 file, split the stream at the 0 bytes and COBS-decode each message; the message formats are described in the protocol documentation, which i4M provides to customers on request. Simpler: use the i4M Python decoder, which delivers pandas DataFrames that can be saved as CSV, Parquet etc.
Can I order the board without USB connector? Yes, for height-restricted installations. Specify radio mode, channel, maximum node count and node number with the order.